01-9932-00 ViewBladder 10 User Guide Revision 2 15
9.
Storage
When the Probe is not being used, it should be stored in a clean, dry area.
Y
CAUTION
Do not use cables or accessories other than those provided with the Interson USB Ultrasound
Probe, as this may result in increased electromagnetic emissions or decrease immunity to such
emissions.
Do not store the probe in the shipping case. It may become a source of infection.
To prevent damage to the probe, do not store in areas where it might be exposed to:
x
Excessive vibration
x
Excessive dust & dirt
Store the probe under the following ambient conditions:
x
Temperature:
-10°C to 50°C (14°F to 122°F)
x
Relative Humidity:
20% to 80% (no condensation)
x
Atmospheric pressure: 700 hPa to 1060 hPa
10. Transportation
Never carry the probe by the cable. The cable could disconnect from the probe allowing it to drop
and possibly damaging the probe.
Never bend the USB cable in a tight radius. This could result in damage to the cable.
Transport the probe under the following ambient conditions:
x
Temperature:
-10°C to 50°C (14°F to 122°F)
x
Relative Humidity:
20% to 80% (no condensation)
x
Atmospheric pressure: 700 hPa to 1060 hPa
When transporting the probe to a different field location or being returned for repair and/or
maintenance, use the disinfected carrying case or enclosure that the probe was originally
packaged in. Call Interson for an RMA number before returning a probe for evaluation and
possible repair. If the original package is not available, pack in such a way that the probe is
protected.