International Gas Detectors Ltd            T: +44(0)161 483 1415

 

V1.0                  

            Email:

[email protected]

                          

 

               

- 14 - 

Inaccurate 

indication  

Sensor is overdue 

service and 

calibration 

Please contact IGD for 

service and calibration  

No bump test 

performed  

Bump test and zero the 

sensor. 

Fault  indication 

of time 

Battery voltage is 

used up 

Please charge it and reset 

time 

Strong 

electromagnetism 

disturbance 

Please reset time 

Zero calibration 

is unavailable 

Too much zero drift of 

gas sensor 

Please calibrate or 

replace the gas sensor 

Minus gas level 

displayed  

Gas sensor drift  

Calibrate zero point 

Sensor fault 

indication  

Sensor fault  

Please contact the 

manufacturer or dealer to 

replace the gas senor 

6. Notices          

6.1 Do not drop from height or expose to vibration. 
6.2 The detector may not work properly if exposed to high 
concentration of gas. 
6.3 To avoid incorrect result or possible damage to the detector, 
please operate and handle the detector in  accordance with the 
manual. 
6.4 The detector should not  be  stored or used under the 
circumstance with caustic gas (such as Cl

2

), nor under other 

rugged  circumstances, including excessive high or low 
temperature, high humidity, electromagnetic field. 
6.5 If there is dust on the surface of the detector after a long-term 
use, please clean it lightly with clean soft cloth. The surface may be 
scraped or destroyed with caustic solvent or abrasive materials. 

 

6.6 To assure the testing accuracy, the detector should be 

Summary of Contents for POR-MULT

Page 1: ...V1 0 Document Reference POR MULT Portable Multi Gas Detector User Manual International Gas Detectors Ltd ISO9001 2015...

Page 2: ...ctors as an extra level of safety 1 Main features Natural diffusion sampling method and high sensitivity sensor with a high sensitivity and repeatability 32 bit built in MCU high reliability and self...

Page 3: ...EL 1 LEL H2S 0 10ppm 5ppm 10ppm 1ppm CO 0 100pm 20ppm 55ppm 1ppm O2 0 30 vol 19 5 vol 23 vol 0 1 vol If other gases are required please contact IGD email below Gas sampling method Natural diffusion De...

Page 4: ...ons 130mm 67mm 30mm L W H 4 Operation Function 4 1 Turn on Press the button for 3s and then release it After the buzzer gives short sound once the detector is turned on After 30 seconds it enters dete...

Page 5: ...nger displays any information and will shut down Fig 2 4 3 Button function Detector in the normal detection state as shown in figure 1 Press any key to operate the back light user can set back light t...

Page 6: ...ashes synchronously and the gas concentration value on the screen changes the colour to yellow In addition to this the back light is illuminated and the vibrator is also activated to indicate low alar...

Page 7: ...plus the change of the gas concentration value is displayed on the screen The text colour turns red the back light and the vibrator are also activated Press key to mute but still display alarm inform...

Page 8: ...rst line shows the time alarm signs it occurs when there is alarm yellow colour with low alarm red colour with high alarm the buzzer instructions normal mute battery percentage battery power date and...

Page 9: ...menu move the cursor to ESC then press to return to the normal monitoring display as shown in Figure 6 Fig 6 4 4 1 Alarm record and alarm curve a Alarm Record Function On the detectors the main interf...

Page 10: ...0 b Curve display function On the main interface press the to enter the curve display interface use the button to view each sensor gas concentration curve On the portable detector each sensor has a cu...

Page 11: ...ings time setting function The menu function is as follows Description Operation Return Press key to move the cursor to the return function press to return to the main interface Setting Sensor paramet...

Page 12: ...alarm and low alarm As an example with Ex as shown in Figure 10 use the key to select the gas type Ex press the to enter Ex gas parameter settings as shown in Figure 11 use the key to modify the param...

Page 13: ...the user should not operate this Fig 15 4 Language In the menu function interface select the language press the key you can see the interface shown in Figure 15 through the key to move the cursor to...

Page 14: ...nd AC power source the detector will be turned on automatically When the battery mark on the screen is full and does not change any more it means the charging is completed Disconnect when charging is...

Page 15: ...lease contact the manufacturer or dealer to replace the gas senor 6 Notices 6 1 Do not drop from height or expose to vibration 6 2 The detector may not work properly if exposed to high concentration o...

Page 16: ...before and after use 6 7 Please dispose of the used Lithium batteries to the appointed places or send to our company Do not discard them into the general waste 7 Standard accessories supplied Metal Ca...

Page 17: ...Email sales internationalgasdetectors com 16 International Gas Detectors Ltd Triton House Crosby Street Stockport Manchester United Kingdom SK2 6SH Tel 44 0 161 483 1415 E mail sales internationalgas...

Reviews: