International Gas Detectors Ltd T: +44(0)161 483 1415
V1.0
Email:
- 12 -
Fig. 13
Fig. 14
parameter value, as shown in Figure 14,
modify the cursor to move to the last press the
▼
key to return to the previous menu.
Warning: 1. zero calibration, please ensure you
use Nitrogen/Instrument Air, otherwise will
affect the accuracy of gas detector.
2. the detector has been calibrated with
standard gas in factory, the user should not
operate this.
Fig.15
4. Language
In the menu function interface, select the
language, press the "
" key, you can see the
interface shown in Figure 15, through the
▲▼
key to move the cursor to select language.
Fig.16
5. Information
In the menu function interface, select the
information, press the "
" key, you can see
the product information shown in Figure 16,
including software version, factory date.
Fig.17
6. Back light
In the menu function interface, select the light,
enter the back-light settings interface as shown
in Figure 17, press the "
" key to move the
cursor, use the
▲▼
key to set the seconds, the
maximum can be set to 120s.