International Gas Detectors Ltd T: +44(0)161 483 1415
V1.0
Email:
- 13 -
Fig. 18
7. Time set
In the menu function interface, select the time,
enter it as shown in the figure 18, press the "
" key to move the cursor, use the
▲▼
key
to set the current system time, press the
▼
key
return to the main menu.
4.5. Charging
Please charge the detector when it shows low battery, or the
detector cannot be turned on due to low battery. Before charging,
please turn off the detector. After you connect the charger correctly
between the detector and AC power source, the detector will be
turned on automatically. When the battery mark on the screen is full
and does not change any more, it means the charging is
completed. Disconnect when charging is complete.
Warning:
Please do not charge it when the detector is
working to avoid potential damage.
Note: Make sure full charge for at least once within 3 months
since production date.
5. Possible fault and corresponding solution
Possible fault
Possible reason
Corresponding solution
The detector
cannot be turned
on
Too low battery
Please charge
The sensor failure
Please contact IGD
Fault of electric circuit Please contact IGD
No response to
the gas
Warm up is not
finished
Wait till warm up is
finished
Fault of electric circuit Please contact IGD