Document No. 715-7000
Revision E
Page 53 of 63
AERONOX
TROUBLESHOOTING GUIDE
FRONT PANEL - OPERATIONAL MESSAGES
BUTTON
PRESSED
PURPOSE
MESSAGE
Alarm
Silence NO
Silences audible alarm for 1
minute
ALARM SILENCE
Visual alarm persists until rectified
Alarm
Silence NO
2
Silences audible alarm for 1
minute
ALARM SILENCE
Visual alarm persists until rectified
ZERO NO
Sets calibration baseline for
selected sensor
CAL NO
, after 5 seconds,
DONE!
ZERO NO
2
Sets calibration baseline for
selected sensor
CAL NO
2
, after 5 seconds,
DONE!
ZERO NO &
ZERO NO
2
PRESSED
TOGETHER
Displays the total time in
hours that both sensors have
been in service
SENSOR TIME
, after 5 seconds, time
in hours that both sensors have been
in service
Visual LED
Alerts user that the sample
line has become obstructed
SAMPLE OBSTRUCTION