Document No. 715-7000
Revision E
Page 33 of 63
Diagram # 9
1.
Quick Connect Gas Inlet Female high pressure gas inlet, receives male
connector on SS hose from delivery regulator
2.
Sample Gas Inlet
Male luer connector. Hydrophobic filter from
sample line (P/N 415-0004) attaches here
3.
Delivery Gas Outlet
Female luer connector. Delivery line (P/N
415-0001) attaches here
CAUTION!
Luer fittings are common to both the AeroNOx
and to most
intravenous systems. Careful attention must be paid when
connecting and disconnecting luer connections to avoid inadvertent
administration of NO gas into a patient I. V. system.
Quick Connect Gas Fittings
The AeroNOx
was designed with quick connect gas fittings for quick and easy connection from
the AeroNOx
to the delivery gas regulators. This can be utilized either when changing from an
empty to a full cylinder, or when switching from a bedside to a transport cylinder and vice versa.
CAUTION!
Ensure the regulator is purged prior to connection to the patient
circuit. See “Purge procedures to follow when changing tanks
and/or regulators” in SECTION II.
To connect the fittings
, insert the male quick connect on the SS hose into the female quick
connect located on the AeroNOx
unit and on the delivery regulators and push them together
until an audible “click” is heard and the two connections stay together.