
Document No. 715-7000
Revision E
Page 12 of 63
Do not substitute AeroNOx
Bagger components! The AeroNOx
Bagger has been
designed and tested for patient safety with the components included in the current
configuration.
The
AeroNOx
Bagger is designed for use only with the AeroNOx
Delivery System. Do
not attempt to use the AeroNOx
with any other manual resuscitator. Do not attempt to
use the AeroNOx
Bagger with any other delivery system.
A backup system must always be available in the event that the primary system should
fail. For instructions for use of a Back-Up System in the event of a failure of the
AeroNOx
system, please see
SECTION III
, Back-Up system.
The
AeroNOx
Bagger is intended to be connected directly to the patient’s endotracheal
tube. Do not insert any additional tubing between the AeroNOx
Bagger and the
endotracheal tube.
The
AeroNOx
Bagger is for single patient use only. Do not reprocess.
Persons using this device should be trained and experienced in the use of this device to
assure affective administration of nitric oxide, and to avoid injury to the patient or to
others resulting from inhalation of excess nitric oxide, nitrogen dioxide, or other reaction
products.
Persons using this device who may be particularly sensitive to nitric oxide, or who may be
exposed to these gases for prolonged periods as a result of the use of this device, should
be aware that the gases exiting the AeroNOx
Bagger are not scavenged.
1.
Connect the large end of the bagger gas supply tubing to an oxygen source. Ensure that
the small end of this tubing is connected to the NO / O
2
gas inlet port on the bagger.
2. Connect
AeroNOx
Bagger gas delivery line to the AeroNOx
delivery gas port.
3.
Connect male end of the sample line (P/N 415-0004) to the sampling port on the bagger
gas supply tubing and the female end to the AeroNOx
sampling port.
4. Set
O
2
flow on the O
2
flow meter to exactly 10 LPM. Let the oxygen run for at least 20
seconds to flush out any residual NO or NO
2
left in the AeroNOx
Bagger.
5.
Turn on NO delivery gas source to 0.25 LPM. If you have correctly set your O
2
flow at
exactly 10 LPM, the delivered NO concentration will be ~20 ppm and delivered NO
2
values should not exceed 0.2 ppm.
Do not exceed 20 ppm during manual ventilation!
6.
Connect test lung and set overflow valve to desired positive end expiratory pressure
(PEEP).