Zero Calibration:
1.
Press and hold the Menu button to enter the setup menu. A beep will sound as the
setup menu is activated.
2.
Using the → (Menu) button, scroll until Calibration is displayed on the screen. Press
the
↓
(Pk/Val) button to enter the calibration sub-menus.
3.
Using the → (Menu) button, scroll until Set
Zero Point
is displayed on the screen.
4.
At this point,
t
he Zero Calibration is ready to begin. Be sure the load cell is connected
and in an unloaded state. If using a simulator, be sure that the simulator is set to
0mV/V. Press the ↓ (Pk/Val) button to initiate the Zero Calibration. Dashed lines will
be displayed at the bottom right of the screen to indicate that the 9825 is capturing
the zero point.
Key-In Calibration:
The Key-In calibration method is typically only used in emergency cases when the
indicator cannot receive a Live calibration. The Key-In method uses a single point to
establish the span of the load cell. It ignores load cell nonlinearity and any asymmetry
between the opposing loading modes.
To execute a Key-In calibration, follow the steps below
:
1.
Press and hold the Menu button to enter the setup menu. A beep will sound as the
setup menu is activated.
2.
Using the → (Menu) button, scroll until Calibration is displayed on the screen. Press
the
↓
(Pk/Val) button to enter the calibration sub-menus.
3. The Live or Key-In sub-menu is the first Calibration sub-menu and should be
displayed on the screen. Press
the
←
(
Reset
) button
to change the flashing value
from Live to Key-In. Press the Menu (Enter) button to save this setting.
4. Press the → (Menu) button to change the sub-menu to Rated Output. Press the ↓
(Pk/Val) button to enter the Rated Output sub-menu.
5. Enter the sensitivity of the load cell into the numerical field. This is typically the mV/V
output of the load cell at its rated capacity. Press the Menu (Enter) button to save this
value.
6. Press the → (Menu) button to change the sub-menu to Sensor Capacity. Press the ↓
(Pk/Val) button to enter the Sensor Capacity sub-menu.
7. Enter the rated capacity of the load cell into the numerical field. Press the Menu
(Enter) button to save this value.
8. Press the → (Menu) button to change the sub-menu to Set Zero Point. User should
preform a Zero Calibration as detailed above.