Interface 9825 Installation & User Manual Download Page 11

System Configuration

 

Menu 

Tree

:

Menu 

Sub-Menu 

Description 

Default 

Options 

Data 

Capture 

Sampling Rate 

Number of Samples 

per second. 

100Hz 

30, 40, 50, 60, 75, 80, 100, 120, 

150, 170, 200, 240, 300, 400, 

600, 1200Hz 

FIR Filter (Finite 

Impulse 

Response) 

Reduces the 

influence of nearby 

electrical or 

mechanical noise 

sources. 

O

n

 

Off, On 

SMA Filter 

(Simple Moving 

Average) 

Smooths signal by 

averaging samples 

over a given span. 

Integer values from 1 to 100 

Units 

Cal Basic Unit 

Select engineering 

units displayed 

lbf 

lbf, ozf, N, kN, g, kg, UNIT_NONE 

Nominal Capacity  Sets display output 

range 

100,000 

Integer values from 1 to 100,000 

Disp Resolution 

Set decimal 

placement and 

increments 

1:100,000  Menu options are based on 

Nominal Capacity value 

Menu 

Description

:

Warning:

Do not access the Advanced Menu unless instructed to do so by a qualified technician.

Interface Inc. www.interfaceforce.com

8

9825 Installation & User Manual

Summary of Contents for 9825

Page 1: ...9825 Installation User Manual Installation User Manual 9825...

Page 2: ...instructions Warning Warning means potential danger Example property damage personal Injury or even death Pre Installation Warnings Warning This device must be installed and connected by a profession...

Page 3: ...USB Interface Communication Analog Output Interface I O Control Applications Setpoint Application Alarm Application Indicator Info Software Version Appendix Appendix 1 Command Output Format 1 Continuo...

Page 4: ...is 20 C to 65 C 4 F to 149 F working environment temperature is 10 C to 104 F 14 F to 104 F relative humidity no more than 95 Non Condensing The 9825 digital indicator uses panel installation which re...

Page 5: ...load cell signal connection This indicator provides a 4 5 volt DC excitation voltage to the load cell s The voltage difference between SIG and SIG is about 0 9mV when connected to a load cell with a 2...

Page 6: ...Input Terminal Pin Assignment Four Wire Analog Load Cell or Junction Box Connection Six Wire Analog Load Cell or Junction Box Connection Interface Inc www interfaceforce com 3 9825 Installation User...

Page 7: ...tput for current output 4 20mA 0 24mA or voltage output 0 10V 0 5V Please note that voltage and current outputs cannot be used at the same time We suggest using a PLC or PC to monitor the analog outpu...

Page 8: ...l can be connected to the positive or negative of the power supply The maximum power output of each relay is 90W 5A Output control interfaces and load connection diagram Output control interfaces and...

Page 9: ...ce value will be displayed Display Details The 9825 utilizes a 128 x 32 dot OLED display with adjustable LED back light The table below summarizes the display annunciators Basic Force Display Force Di...

Page 10: ...ay mode PK Val Function Cycle among real time peak and valley display modes When in the setup menu Function Enter sub menu Decrease a value when used as a directional key Reset When in display mode Re...

Page 11: ...mooths signal by averaging samples over a given span 1 Integer values from 1 to 100 Units Cal Basic Unit Select engineering units displayed lbf lbf ozf N kN g kg UNIT_NONE Nominal Capacity Sets displa...

Page 12: ...8 bit None 8 bit Even 7 bit Even 7 bit Odd Port Mode Set port mode Demand Demand Continuous Protocol Set Protocol see protocol descriptions in the appendix Condec Condec ASCII Analog Out Output Type S...

Page 13: ...he screen Press the Pk Val button to initiate the calibration process 4 The term Fixture will show on the screen At this point the load cell should be set into its fixturing but with no additional cal...

Page 14: ...follow the steps below 1 Press and hold the Menu button to enter the setup menu A beep will sound as the setup menu is activated 2 Using the Menu button scroll until Calibration is displayed on the s...

Page 15: ...t of the 9825 should be connected to whatever instrument will be accepting and measuring the analog signal 1 In the Analog Output setup menu scroll to Fine Tune and press the Pk Val button to initiate...

Page 16: ...s less than the value of Input Point 4 but greater than the value of Input Point 3 The symbol will show on the display The OUT 4 relay will close Otherwise the symbol will show on the display and the...

Page 17: ...relay will open 4 When the load is less than the value of Input ExtraLow The symbol will show on the display The alarm will sound The OUT 4 relay will close The display will send a warning message Oth...

Page 18: ...Format 1 Continuous Mode ASCII In this mode of communication the indicator transmits the data frame continuously The load value in the frame is expressed in ASCII Interface Inc www interfaceforce com...

Page 19: ...device PC will output demand command through the serial ports when the scale is in the normal loading status The demand command format is shown below The serial output data format is as follows Interf...

Page 20: ...ndec Demand Output Demand Commands P Print T Tare Z Zero G Gross N Net Condec Continuous Output Note The starting address 40001 of MODBUS is not suitable for SIEMENS soft Interface Inc www interfacefo...

Page 21: ...C 10 to 90 non condensing at F 10 to 90 non condensing POWER Supply VDC 24 VDC with supplied 120V 60Hz AC DC adapter or 9 36 VDC external supply Power Consumption W 6 RMS 8 Peak MECHANICAL Dimensions...

Page 22: ...freight insurance and proper packaging to prevent breakage in transit Interface warranty does not apply to defects resulting from action of the buyer such as mishandling improper interfacing operatio...

Reviews: