9 STARTING
We have to make sure that the Fluorescent bulb is undamaged. We put faultless bulb gently
in the frame.
Before turning on we have to be sure, that voltage is the same as on the plate on the back of
the instrument.
If voltage does not match, you must not connect the instrument.
With pushing on the switch
4
on 0, you can make sure, that the instrument is off.
Connect network cable of the instrument
5
with well installed and safe electric source.
Turn the network switch 4 with push on I – the instrument is ready.
With switch
7
you choose method of work:
O
continuous activity
I
Activity for 5 minutes
II
Activity for 10 minutes
With push on the start-button
8
the fluorescent bulbs in the polymerisation chamber will be
turned on for requested duration.
Damaged fluorescent bulbs should be immediately changed. The instrument should not
be operating with damaged bulbs.
In case, that one or more fluorescent bulbs do not work, or that fluorescent light does not
turn off after chosen time, use the instructions in the table.
You turn of the instrument by pushing switch 4 on 0.
10 MAINTAINCE
Interlux usually does not need mintaince. For safe work you have to check before every use,
that instrument does not have external damages and that network cable is faultless.
Before every use you should make sure that fluorescent lights are well installed. Broken
lights should not be used.
Repairs of functional parts performs manufacturer Interdent. You can use only original
fluorescent lights, according to rules of manufacturer. Interdent is not responsible for the
damage, made because of using damaged parts or parts, not delivered by Interdent.