background image

 

While repairing and maintaining the instrument must not be connected on wire netting.  

 

Only authorized person is allowed to repair the instrument. Interdent is not responsible for 
the damage, made by non-expertise repairs on the instrument. 

 

After repairs on the electrical parts of the instrument, it should be tested again if it fits local 
regulations. 

 

 

6 SAFETY USAGE 

To prevent stagnation of overheating, there are grids on the upper side of the instrument, 
which have to be free during work. The instrument has to have safe distance from heating 
devices.  

With  damages,  which  do  not  assure  safe  work  anymore,  the  instrument  should  be 
disconnected from wire netting and usage should be prevented, until the instrument is not 
flawless.  

To prevent unwanted polymerisation, you should not put synthetic materials, which can be 
polymerised by light, in front of the instrument.  

 

 

7 SETTING AND STARTING 

Unpacking 

When you are unpacking, you should make sure, that instrument is not damaged. 

In  case  of  damage,  you  should  return  the  instrument  to  manufacturer  with  the  bill  and 
supply order, which includes serial number of the instrument.  

Damages,  caused  by  incorrect  transport,  should  immediately  be  laid  claim  to  authorized 
forwarding agent.  

 

 

8 PLACING 

The instrument should be placed on horizontal, flat and non-sliding working surface, on the 
distance from heating devices. You should be careful, that there is enough room for on-off 
switch and that grids for cooling polymerisation chamber are free.  

Before first turning on, you should wait that the instrument heats on room temperature.  

 

Summary of Contents for INTERLUX

Page 1: ...INTERDENT D O O Opekarni ka cesta 26 3000 Celje Slovenija Tel 386 3 425 62 00 Fax 386 3 425 62 10 Email info interdent cc www interdent cc Instruction for use INTERLUX...

Page 2: ...IONS FOR SAFE WORK 5 5 ELECTRIC OPERATIVE RELIABILITY 5 6 SAFETY USAGE 6 7 SETTING AND STARTING 6 8 PLACING 6 9 STARTING 7 10 MAINTAINCE 7 11 CLEANING AND CARE 8 12 DISTURBANCES AND THEIR RESOLVING 9...

Page 3: ...3 1 VOLUME OF SUPPLY PART DESCRIPTION 1 Casing 2 Polymerisation chamber 3 4 fluorescent lamps 4 Lattice switch 5 Lattice cable 6 Fuse 7 Switch for choosing the method of work 8 Start switch Art t 0715...

Page 4: ...Weight 1 77 kg EU US Voltage 230 V 115 V Frequency 50 Hz 60 Hz Taking away power 36 W 36 W Influence on environment Temperature 5 C 40 C Relative air humidity 90 at 20 C Plate with labels and data wit...

Page 5: ...ions should be taken under consideration INTERLUX is used exclusively for photopolymerisation of lacquers and synthetic plastic materials which are light consolidating Usage for other purposes is not...

Page 6: ...revented until the instrument is not flawless To prevent unwanted polymerisation you should not put synthetic materials which can be polymerised by light in front of the instrument 7 SETTING AND START...

Page 7: ...will be turned on for requested duration Damaged fluorescent bulbs should be immediately changed The instrument should not be operating with damaged bulbs In case that one or more fluorescent bulbs do...

Page 8: ...RE Before each cleaning the instrument should be turned off You should be careful that there are not any leftovers of materials in the polymerisation chamber It is recommended that you clean the lefto...

Page 9: ...set the fuse reconnect call maintance One or more hose does not shine Hoses are not inserted correclty Check hoses and insert them correctly Damaged hose Insert new hoses Polymerisation takes too lon...

Page 10: ...o Opekarni ka cesta 26 SI 3000 CELJE Was developed constructed and made in accordance with following demands EU directive for instruments 98 37 EU EU directive for low voltage 73 23 EWG changed in 93...

Page 11: ...he warranty period Failure will be resolved free of charge within 45 days from the date of failure report The product which will not be repaired within this deadline will be at your request replaced w...