Operating Instructions
4/8-indent crimping tool CO.236.00
Rev.: 2015-10-16
8 / 12
EN
9 Calibrating your tool (CAL) after changing the battery
Info
Mechanical contact at the lower setting gives the reference value for calculating
the current state of wear. This value is permanently stored in the tool's memory,
and cannot be altered. You will need to calibrate the tool against this reference
value every time you change the battery. Keep to the sequence given below.
· Set the display to mm² using the MODE button (6) (see 2. Operation).
· Set the crimper die using the adjustment wheel (8) until gauge supplied with the crimper
touches the dies and you can move it without play.
· Make sure that the gauge measurement to be set is always taken from a larger value,
such as from 2.05 mm down to 2.0 mm gauge.
· Keep the ON/OFF switch (7) pressed and press the MODE button (6) using the gauge.
Keep the MODE button (6) pressed for at least five seconds.
· Release the MODE button (6) after five seconds, and then release the ON/OFF switch (7).
· The digital display will automatically show a gauge value of 2.0 mm.
· Your crimper is recalibrated and ready for crimp parameter setting.
NOTE
Always have authorised personnel calibrate your crimper; improper calibration will
lead to bad crimps.
8 Recalibrating the crimper (REC)
· Open the battery compartment (1) upwards.
· Remove the used battery.
· Turn the adjustment wheel down to the lower setting (minus sign turning direction) and
leave it there.
· Insert the new battery. The display (5) will show CAL as a calibration request.
· Set the crimper to 1 mm using the gauge – turn the adjustment wheel (8) until you can
move the gauge between the dies without play as shown in 7.
· Keep the ON/OFF button (7) pressed, and press the MODE button (6) with the gauge.
. Make sure that the gauge measurement is always taken from a larger value, such as from
2.05 mm down to 2.0 mm. This means, that at the beginning of the adjustment
the gauge can be inserted with having some play.
· Keep the MODE button pressed for at least five seconds. Release the MODE button after
five seconds, and then release the ON/OFF switch.
· The digital display will automatically show a gauge value of 2.0 mm.
· Your crimper is calibrated and ready for crimp parameter setting.