35
RE
emissions
CISPR1 1
Class B
The device is suitable for use in
all establishments including
domestic and those directly
connected to the public low voltage
power supply network
that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC
61000-3-2
Not
applicable
Voltage
fluctuations
/flicker
emissions
IEC
61000-3-3
Not
applicable
Guidance and manufacturer's declaration —
electromagnetic immunity
The device is intended for use in the electromagnetic
environment specified below. The customer or the user
should assume that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment — guidance
Electrostatic
discharge
(ESD) IEC
6100-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material the
relative humidity should
be at least 30% tool