![intensity TWIN STIM III Instruction Manual Download Page 15](http://html1.mh-extra.com/html/intensity/twin-stim-iii/twin-stim-iii_instruction-manual_2075144015.webp)
15
3. Specification
3.1 Accessories
NO
DESCRIPCIÓN
CANTIDAD
1
Electrical stimulator device
1 piece
2
2 Electrode Leads
2 piecess
3
3 1.5” x 1.5” Adhesive Electrodes
4 piecess
4
4 9V Alkaline Battery, type 6LR61
1 piece
5
5 Instruction Manual
1 piece
6
6 Carrying case
1 piece
7
7 AC Adapter (optional)
1 piece
3.2 Technical information
Channel
Dual, isolated between channels
Power Supply
9.0 V Alkaline DC-Ix6LR61 battery
Adapter output: 9.OVdc, 800mA(optional)
Operating
Conditions
5°C to 40°C ( 41°F – 104°F ) with a relative
humidity of 30% – 75%, atmospheric pressure from
700 – 1060 hPa.
Storage
Conditions
-10°C to 50°C (14°F to 122°F)with a relative
humidity of 10%-90%, atmospheric pressure from
700 – 1060 hPa
Dimensions
4.5 x 2.55 x 0.9 inches(LxWxH)
Weight
0.28 lbs. (With battery)
Tolerance
There may be a ±5% tolerance of all setting
and ±10% tolerance of output of intensity.
Timer
Adjustable from 1 – 60 minutes or continuous.
Adjusts in 1 minute steps. Treatment time
countdown is automatic.
Electrode
Detection
Function
The amplitude level will be reset to OmA when the
amplitude level is 12mA or greater and an open
circuit at either channel is detected.