This condition is met when the operator pressing the STOP key terminates the charge or
test.
C V
F A I L
This condition is specific to Lead Acid batteries being charged in constant voltage mode. If
the current being applied to the battery does not fall below the current limit value within
the current voltage changeover time then this condition is met and the battery is
considered faulty.
B A T T R E V
This condition will terminate and Tests or Charges if the detects reverse
battery connection.
T > 2 4 H
This condition is met if on a 0.1C or 0.2C charge the switch over voltage is not reached.
This prevents the charger continually charging the battery in case a fault condition has
occurred.
P S U F A I L
This condition is met if on charge one or both of the power supplies drop out. This
condition can occur if the thermal trips fitted to the transformers fail or if there is an
electrical failure. Allow the unit to cool down before starting another test. If this
condition is shown upon starting a test then one or both of the power supplied have
failed and service of the unit will be required.
? E R R O R ?
This condition will occur if there is a problem with the demand circuit. This can occur if
battery is being capacity tested at voltages & currents outside the units capabilities.
V O L T S
This condition will occur in constant voltage charge mode, if the battery voltage is
approximately 8.0V or more above the constant voltage set point.