Technical data
6.1
Specifications
Device
Type
Safety Bunsen burner
Dimensions (W x D x H)
100 x 150 x 60 mm without burner head
Weight
1.3 kg
Electricity supply
Mains adapter
Input: 100–240 VAC, 50/60 Hz
Device
Input: 8.6–9.4 VDC, 0.5 A, max. 5 W
Materials
Metallic Housing: zinc
Burner head: stainless steel, brass
Display: glass
Base: PBT (polybutylene terephthalate) /
PC (polycarbonate)
Operating conditions
Operation
15–35 °C, 15–80% RH (not condensing)
Storage
-10–60 °C, 5–80% RH (not condensing)
Nominal thermal rating
Propane/butane gas (liquid gas) 2.0 kW
Natural gas E (~94% methane)
1.0 kW
Natural gas LL (~81% methane) 0.8 kW
Nominal gas input pressure
Propane/butane gas (liquid gas) 50 mbar
Natural gas E, LL
20 mbar