FIREBOY eco - Operating instructions V07
Installation
www.integra-biosciences.com
15
Engl
ish
3.3
Connecting the foot or benchtop switch
Insert the plug of the foot or benchtop switch in the socket on the reverse
side of the device (see “Rear view” on page 10).
The switch is now connected.
3.4
Removing and adding the burner head
To remove the burner head, pull it out vertically upward.
To install the burner head, place it on the device with the 3 contact pins and
push it down vertically until it clicks into position.
The burner head is now reinstalled.
C
AUTION
Never carry the device by the burner head.
3.5
Long burner head
If the FIREBOY eco is used for continuous operation, the optionally available
long burner head must be used. For installation see previous section.
C
AUTION
The minimum distance between the tip of the burner head and the object to be
heated, e. g. cover plate of quadruped, must be at least 60 mm.
The duration of continuous operation should not exceed 60 min.