FIREBOY eco - Operating instructions V07
Installation
www.integra-biosciences.com
11
Engl
ish
3
Installation
Purpo-
se
This chapter defines the requirements for the gas connection and the electricity
supply of FIREBOY eco.
3.1
Connecting FIREBOY eco to the electricity supply
Insert the mains plug in the socket on the reverse side of the device (see
“Rear view” on page 10) and connect to the power source.
3.2
Connecting FIREBOY eco to the gas supply
FIREBOY eco is appropriate for use with natural gas (gas mixture, >81%
methane, further alkanes) and commercial butane/propane gas (mixture 65%–
10% butane, 20%–40% propane).
3.2.1 Fitting the nozzle for the gas in use
The P60 nozzle for butane/propane gas is fitted ex-factory for FIREBOY eco. If
necessary, the nozzle must be fitted for the gas in use.
The nozzles are marked on the lower surface with the appropriate designation.
C
AUTION
FIREBOY eco may only be operated with the nozzle that corresponds to the
gas type in use. Only use the nozzles supplied. Prior to every nozzle
installation, inspect the O-ring in the brass screw for damage and if necessary
replace it.
Type of gas
Gas input pressure
Inscription nozzle
Natural gas (methane)
20 mbar
N80
Butane/propane gas
50 mbar
P60