Intec PolyGard 2 DT6 User Manual Download Page 7

4.4 Calibration 

The 

STL6

 service tool or 

DPT6 EasyConfig 

Software are available for convenient on-site calibration. 

There is also the possibility to exchange the SC  against  a  calibrated  SC  on site. The used SC  can then be 
calibrated directly in the office or at 

INTEC Controls

 and then reused.  

Prior to calibration the Sensor Cartridge  must be supplied with power voltage without 
interruption for warm-up and stabilisation. The warm-up time depends on the sensor element 
and is shown in the following table for Example: See also User Manual for Sensor Cartridge. 

Sensor Cartridge 

Warm-up for 
calibration 
(h) 

Stabilization time 
to specification 
(min) 

Flow rate 
(ml/min) 

CO 

24 

60 

150 

NO

2

 

24 

60 

500 

LPG (C

3

H

4

24 

60 

150 

Table Calibration 

Please observe proper handling procedures for compressed gas and test gas bottles (regulations 
TRGS 220)!  

Test gas can be toxic, so never inhale it!  
Symptoms: Dizziness, headache and nausea.  
Procedure if exposed: Remove victim to fresh air, seek medical attention. 

4.5 Calibration with 

DPT6

 

EasyConfig 

Software

Connect 

DPT6 EasyConfig Software

 Interface to the 

DT6 

Basic Sensor Board.

Prior to calibration you have  to activate  the  mode  “Special Mode” at  the BSB, only  then the
calibration menu is enabled. During the special mode the BSB doesn’t issue alerts.

Select the Sensor Cartridge to be calibrated by selecting the gas type.

Zero calibration 

The current zero offset and the offset value of the first calibration is read with "Read".

Slide calibration adapter carefully onto the Sensor Cartridge.

Apply synthetic air (flow rate according to the table "calibration", 1 bar ± 10%) to the Sensor
Cartridge.

When the value is stable, the new zero offset factor is calculated with "Calibration".

The new offset factor is checked for plausibility and stored in the buffer memory. The current measured 
value is output with the new offset factor and the offset display is updated.  

With "Save" the new offset factor is written in the SC memory, only then the Zero calibration has
been successfully completed. If you exit the menu without pressing "Save", the original offset data

for the measured value calculation will continue to be used.

With a zero reading> 10% of measuring  range during the zero calibration, zero calibration is 
not possible.  

INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com

Specifications subject to change without notice. | GASB2_03_E_0220 | USA 200306 | Page 7 of 15

DT6 – UserManual

Summary of Contents for PolyGard 2 DT6

Page 1: ...Transmitters Serial No SB 03 User Manual Version February 2020 en ref optional remote sensor DR6 INTEC Controls 12700 Stowe Drive Suite 100 Poway CA 92064 Ph 858 578 7887 888 GO INTEC inteccontrols c...

Page 2: ...5 Inspection and Service 9 5 1 Inspection 9 5 2 Service and Calibration 9 6 Troubleshooting 10 6 1 DT6 Basic Sensor Board 10 6 2 SC2 Sensor Cartridge Messages at the Tool Controller 10 7 Technical Dat...

Page 3: ...local bus level is done via the STL6 Service Tool or with the DPT6 EasyConfig Software which is directly connected to the BSB The cable topology for the RS 485 field bus can be taken from the Guideli...

Page 4: ...eight depends on the relative density of the gas type to be monitored Choose mounting location of the sensor according to the local regulations Consider ventilation conditions Do not mount the sensor...

Page 5: ...SB connect cable according to connection diagram fig 1 Replug terminal carefully on the BSB at X4 For remote sensor Insert and connect local bus cable at the basic and remote sensor board Remove termi...

Page 6: ...as type and defining the measurement range the SC is connected to the input Up to two different SC can be connected at the DT6 with a third to an attached remote DR6 The physical position of the SC do...

Page 7: ...on you have to activate the mode Special Mode at the BSB only then the calibration menu is enabled During the special mode the BSB doesn t issue alerts Select the Sensor Cartridge to be calibrated by...

Page 8: ...g the SC connector which leads to an immediate triggering of the error message Disconnect the SC connector from the BSB or the RSB error message will be activated Loosen the locknut Remove used SC Tak...

Page 9: ...be documented The date for the next maintenance has to be affixed to the sensor 5 1 Inspection Gas sensors should be controlled regularly by a competent person according to EN 45544 4 The following ha...

Page 10: ...h with registered ones Check SC data registration data for conformity Message at the Tool Controller 24 V DC voltage range 5 V DC voltage range Temp range WatchDog triggered Internal error Replace BSB...

Page 11: ...B x D 94 x 130 x 57 mm 3 7 x 5 1 x 2 2 in Weight 0 3 kg 0 7 lb Protection class delivery status NEMA 4X IP 65 Installation Wall mounting Wire connection Field bus Local bus for SC Screw type terminal...

Page 12: ...oard Local Bus only for Remote Sensor Board Field Bus 1 2 3 4 0 VDC 24 VDC Bus_A Bus_B Bus_B Bus_A 24 VDC 0 VDC Analog Output 4 20 mA 0 VDC Field Bus 1 2 3 500 Remove resistor if device is connected X...

Page 13: ...rd with SC X 2 X 2 X 2 X 2 X 2 X2 X 3 X 3 X 3 SB_03 SB_03 SB_03 SB_03 SB_03 SB_03 X4 4 3 2 1 X4 4 3 2 1 X4 4 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 Basic Sensor Board Basic Sensor...

Page 14: ...Mounting Mounting Mounting Mounting 115 mm 130 mm 180 mm D 4 mm INTEC Controls 12700 Stowe Drive Suite 100 Poway CA 92064 Ph 858 578 7887 888 GO INTEC inteccontrols com Specifications subject to chan...

Page 15: ...de ANSI NFPA70 The equipotential bonding required also e g secondary potential to earth or grounding measures must be carried out in accordance with the respective project requirements It is important...

Reviews: