53
Residual risks
Product Support:
In addition, if the gap between grips and fixtures is less than 50 mm:
•
Use a specimen insertion tool and other appropriate equipment and practices to
keep fingers out of the space between the fixtures.
In addition, if the gap between grips and fixtures is less than 25 mm:
•
Use a specimen insertion tool to keep fingers out of the space between the fixtures.
•
Use an interlocked shield to limit or disallow motion when the shield door is open.
When the space is less than 25mm, the crush hazard is very significant, so this is
the best option.