Chapter 3: Additional System Details
50
M47-17036-EN
11
•
The HPS shuts down
•
Software posts a motor temperature
error message
Oil level has dropped below the
allowable level
Perform checks and/or corrections for
an oil leak and then add oil to the
reservoir. Refer to procedures provided
in Chapter 4 of the System Operating
Instructions (supplied separately).
Oil temperature exceeds
temperature switch setpoint,
possibly due to:
•
Fans not operating properly
•
Check for and correct fan operation
•
Environmental conditions
exceed requirements
provided in Chapter 2 of the
System Operating
Instructions (supplied
separately)
•
Correct environmental conditions
Oil temperature must drop below the maximum temperature stated in
to reset the switch. Once the
switch has reset, the HPS can be started.
12
•
For frames with a J3-style
crosshead, the hydraulic cylinder will
not jog or a test will not start
The locks control valve is open
Refer to information provided in the
System Operating Instructions (supplied
separately)
The HPS was not able to build
enough pressure to clamp the
crosshead lock cylinders before
the
Jog
controls were used
Refer to information provided in the
System Operating Instructions (supplied
separately)
13
•
Pump Start
button is lit but HPS is
not running
The disconnect switch not
ON ( | ) or customer’s power
supply not ON
•
Turn the disconnect switch to ON ( | )
•
Turn the customer’s power supply to
ON
14
•
The software posts an error
message like the one shown in
•
EFI cable could be damaged
or have loose connections
•
Communication problem
between EFI and 59 Series
control unit or EFI and
software
•
Incorrect version of
controlling software is in use
•
Check the cable for damage and
check that connections are secure
•
Check that the correct Ethernet
crossover cable is being used
•
Check IP settings, they should match
those shown in
•
Verify that the software version is
8.2a or above for Partner, or 2.15 or
above for Bluehill
•
If none of the above correct the
problem, then contact your local
Instron Services department as
directed on page
Table 9. Troubleshooting the KPX system. (Continued)
No.
Possible Indications
Possible Problem
Solution