Chapter 2: Installation Notes
20
M47-17036-EN
Electrical interconnections
Before making electrical interconnections, all system components (including the 59 Series control unit,
pendant station and computer system) should be unpacked and in place at the installation site. Also, the
computer system components can and should be interconnected as outlined by the computer
manufacturer. Check the electrical nameplate of each computer component and connect to an
appropriate power source.
,
as you step through the interconnection process.
Electrical cables that connect to components on the frame are routed through the flexible track that is
mounted at the rear of the frame, so these connections are made at the bottom of the flexible track.
Table 2. Electrical connections required.
Cable
Connect From
Connect To
Ethernet crossover cable, male RJ45
Auxiliary Ethernet connection on
rear of computer
Ethernet connection on front of
59 Series control unit (see
)
User control panel cable, female 7-pin
User control panel (typically
hardwired)
Connection N on rear of 59 Series
control unit
HPS control box cable, male 15-pin
HPS control box (typically
hardwired)
Connection M on rear of 59 Series
control unit
HPS emergency stop cable, male 6-pin
HPS control box (typically
hardwired)
Connection F on HPS connection
panel
HPS cable, male 14-pin
Connection A on rear of 59 Series
control unit
Connection A on HPS connection
panel
Frame cable, male 10-pin
Junction box on frame (typically
hardwired)
Connection R on rear of 59 Series
control unit
Load transducer cable, male 25-pin
LOAD connection on front of
59 Series control unit
Load transducer on frame
Position cable, male 9-pin
Connection H on rear of 59 Series
control unit
Position encoder on frame
1
For systems that include an optional
enclosure
Interlock cable
Junction box on frame - usually
needs to be hardwired
Enclosure interlocks - typically
routed along inside of enclosure to
its exit point
Ground cable
Ground stud on rear of 59 Series
control unit
Ground stud on frame
Power cable for 59 Series control unit
Power module on rear of 59 Series
control unit
Appropriate outlet or power strip
HPS main power cable
Connection D on HPS connection
panel
Customer power supply of
appropriate rating
For systems that include an optional AVE
video extensometer
Coaxial cable, male BNC
SYNC
(BNC) connection on front
of 59 Series control unit (see
)
BNC connection on AVE
extensometer
For systems that include the optional
Versachannel or Multichannel
Coaxial cable, male BNC
SYNC
(BNC) connection on front
of 59 Series control unit (see
)
BNC connection on the
Versachannel or Multichannel box