9.0 

CALIBRATION

  

Remove all weight from the pan before beginning.  

 

When in the normal weighing mode, hold

  [Zero] 

and press the

 [Tare]

 key to 

enter the parameter menu as shown in section 9.0.. 

 

The display will  show the revision number in the 

“Weight” 

display ,The

 “Unit 

Weight”

  display  will  show

  "-  - - -     "

  requesting  the  password  number  to  be 

entered.  

 

The  default  password  is 

"0000" 

but  other  numbers  can  be  set  using  the 

parameter menus. Press the 

[0]

 key four times.  The display will jump to the 

first  parameter.     

"FILt ?" 

with  the  Unit  weight  window  showing  the  current 

filter setting. 

 

Press 

[M+/

↵↵

] 8 times 

until the display shows

 

““““CALI”.   

Note that the display will 

also  show 

““““CALI” “Unit ?”  “   xx”   

for  selecting  the  unit  of  weight  to  be  used 

when  calibrating  after  6  presses.      Ignore  this  function  unless  yu  want  to 
change the weighing unit during calibration. 

 

Press the numeric keys to enter the calibrate mass value.  For kilograms and 

pounds enter the value as a number, for example “3” for 3 kilograms.  For 

grams enter the weight in grams,  i.e. “1000” grams.  

 

Place the calibration mass on the weighing platform.  

 The suggested calibrate values for all model : 

 
 
 
 
 
 
 

However any whole value of weight can be used by entering the correct value. 

 

 

Press the 

[M+/

↵↵

]  

key

,

  

 

The display will show “

PASS”

,  

 

Press the 

[M+/

↵↵

]  

key to return to normal weighing. 

 

If the calibration is  outside of the previous calibration value by more than 

5% ,it will show 

“FAIL L” 

or 

“FAIL H”

  indicating the new calibration weight 

was too low(FAIL L)  or too high (FAIL H) in respect to the previous 

calibration value. 

       

 

LBC 3 

LBC 6 

LBC 15 

LBC 30 

2kg or 3kg 

5kg or 6kg 

10kg or 15kg 

20kg or 30kg 

5lb or 6lb 

10lb or 12lb 

20lb or 25lb 

40lb or 50lb 

Summary of Contents for IBC-15

Page 1: ...Inscale Ltd 2011 IBC SERIES Pn 700660168 RevisionC August 2011...

Page 2: ...Inscale Ltd 2011 Easy Reference Model name of the scale Serial number of the unit Software revision number Displayed when power is first turned on Date of Purchase Name of the supplier and place...

Page 3: ...SPLAY 8 5 2 UNIT WEIGHT DISPLAY 8 5 3 TOTAL DISPLAY 8 6 0 KEY DESCRIPTIONS 9 7 0 WEIGHING 9 7 1 BEFORE WEIGHING 9 7 2 WEIGHING 9 7 3 ZEROING THE DISPLAY 9 7 4 TARING 10 7 5 PARTS COUNTING 10 7 6 ACCUM...

Page 4: ...Inscale Ltd 2011 p a g e 2...

Page 5: ...operation Four models 3kg X 0 1g 6kg X0 2g 15kg X 0 5g 30kg X 1g Large LCD with backlight 3 weighing units g kg lb Indicators for Stable zero net weight and low battery functions Stainless steel pan...

Page 6: ...Inscale Ltd 2011 p a g e 4...

Page 7: ...0 2g 0 5g 1g Repeatability 0 1g 0 2g 0 5g 1g Linearity 0 2g 0 4g 1g 2g Units kg g lb Stabilization time 2s 2s 2s 2s Operating temperature 0 C to 40 C 32 F to 104 F Power supply 12VDC 800mA adapter 6V...

Page 8: ...be removed if the scale is not used for a long period of time Do not stack material on the scale when it is not in use Avoid high humidity that might cause condensation and keep away from direct cont...

Page 9: ...ould be adjusted such that the bubble in the spirit level is in the centre of the level and the scale is supported by all four feet Attach the power supply cable to the connector on the right side of...

Page 10: ...ator Zero Stability indicator Stable Indicators for the changed units kg g lb 5 2 UNIT WEIGHT DISPLAY This display will show the unit weight of a sample This value is either input by the user or compu...

Page 11: ...e value from the weight and shows the results This is the net weight Entering a value using the keypad will store that as the tare value Zero Sets the zero point for all subsequent weighing It shows z...

Page 12: ...roduct was to be added to the first one Again only the weight that is added after taring will be displayed 7 5 PARTS COUNTING In order to do parts counting it is necessary to know the average weight o...

Page 13: ...and the quantity will increase If the function ASPL is enabled see section 6 1 the unit weight will be automatically recalculated when an additional mass less than the current count is added If the sc...

Page 14: ...key Press the M key will return to the normal weighing 7 7 BACKLIGHT FOR LCD The backlight of the LCD can be set to be light or dark When in normal weighing mode press the Sample Blight key to change...

Page 15: ...e to be used to determine zero tracking range of 0 5d 1d 2d 4d or Off A larger number means larger range of zero tracking and a worse repeatability the default is 1d CodE Use the keypad to set a new p...

Page 16: ...en calibrating after 6 presses Ignore this function unless yu want to change the weighing unit during calibration Press the numeric keys to enter the calibrate mass value For kilograms and pounds ente...

Page 17: ...Inscale Ltd 2011 p a g e 15 10 0 ERROR MESSAGE Z ERR A D counts are too high or low EEEEE Overload 9d Arrow above Lo Bat Low battery indication...

Page 18: ...Inscale Ltd 2011 p a g e 16...

Page 19: ...in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with...

Page 20: ...listing of all Inscale products visit our website at www inscale co uk Copyright by inscale Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any mea...

Reviews: