The Parameter menu has 9 functions. To set the value for any parameter
press
[Tare]
to decrease or
[Zero]
to increase the parameter and press
[M+/
↵
↵↵
↵
]
to save and move to the next setting
Weight Display
Description
FILt
Set a value for the amount of filtering to be done ranging
from 1 to 7. A larger number means more filtering and a
slower response. The default setting is 3.
StAb
Set a value to be used to determine balance stability, set a
value of 0.5d,1d, 2d, or 4d.
A larger number corresponds to a larger stable zone. Default
is 1.
ZtrC
Set a value to be used to determine zero tracking range of
0.5d,1d,2d,4d or Off, A larger number means larger range of
zero tracking and a worse repeatability. the default is 1d
CodE
Use the keypad to set a new password. The default is
“0000”
bLIg
Setting the backlight
Options:
On
OFF
AUTO (default)
A SPL
In counting mode, this function allows the scale to
recalculate the unit weight automatically after the first unit
weight calibration. Recalculation of the unit weight is
beneficial because errors may occur as the counting
accumulates. The option is on and off.
CALI UNIt
To set the default unit. The option is g,kg,lb.
ZErO
To get the zero AD count.
CALI
See detail in section 10.0
To return to weighing after setting any parameter press the [Total] key to return to normal
weighing.