The Spirohome
®
Clinic should never be used with a charging smart device. Make sure the
smart device is unplugged from its charger before conducting a spirometry test.
Store and use the device as specified in this user manual as alternative methods or
conditions of storage may affect device function and/or accuracy. Use only in specified
environments/conditions (see Operating Environment) to avoid malfunction and/or display of
incorrect results.
All repairs, modifications or reconfigurations must be performed only by the manufacturer. If
the Spirohome
®
Clinic is damaged or malfunctioning, contact the manufacturer or distributor
if purchased from a reseller, directly to avoid incorrect measurements or potential harm. Do
not attempt to repair the device yourself, an opened device casing will terminate the product
warranty.
Please follow all data security warnings and recommendations for the smart device used in
conjunction with the Spirohome
®
Clinic as per its manufacturer’s instructions as the patients’
personal data recorded and stored on the Spirohome
®
Clinic App, may otherwise be at risk.
The user is encouraged to not share Spirohome
®
app account information with unauthorized
parties.
The Spirohome
®
Clinic conforms to EN 60601-1, EN 60601-1-11, EN 60601-1-2 and EN 300
328. As this device operates with RF technology, it must be used as only specified by the
manufacturer, it may avoid interference to radio communications.
4.
MAINTENANCE
Handle the Spirohome
®
Clinic and Spiroway Disposable mouthpiece with care. Do not use
the device or its accessories if they are visibly damaged, particularly if there is damage to
the filters on the mouthpiece.
Store the Spirohome
®
Clinic in dust-, dirt-, and moisture-free conditions. The pouch provided
with the device may be used to keep the device protected. Before each use, always check
that the device is free from contaminants and does not have any visible damage. After use
by each new user, clean and disinfect the device according to the instructions given in this
section.
Proper cleaning and disinfection of your Spirohome
®
Clinic are important for the safe use of
the device. With regular cleaning, the physical build-up of contaminants on the device can be
prevented. A cleaning process must always precede a disinfection process. Disinfection
destroys any pathogens such as bacteria, viruses or other microorganisms that might still be
present on device surfaces after an initial cleaning process. Regular, thorough cleaning and
disinfection of the device protect both handlers and users of the device from the potential
transmission of infections resulting from contact with the device. Handlers and users of the
device should be sure to wash hands with soap before and after each use of the device.
NOTE:
One ‘use’ of the spirometry session is defined as one complete spirometric testing
session (can include up to 8 individual successive spirometry tests).
First Pub. Date: 24.09.2018
R.9-1 / 10.03.2020