spiro
home
®
│Clinic
User Manual
24 / 33
4.2.
CLEANING AND DISINFECTION
Important!
: The Spirohome
®
Clinic must be cleaned and then disinfected between each new
patient.
1. Wash Hands
Before beginning the cleaning procedure, wash
hands thoroughly with soap and water.
2. Perform Cleaning
Remove
the
Spiroway
Disposable
mouthpiece
attached to the Spirohome
®
Clinic body by pushing the
notch, refraining from making contact with the
mouth-contacting part of the mouthpiece. Using a
high-level disinfectant (Sodium Hypochlorite <0.6%,
Sodium hydroxide <0.02%, Sodium Chloride <4%)
wipe, wipe (for at least 30 seconds) all accessible
surfaces of the device to remove all visible
contaminants as shown below. Please be extra careful
and gentle when cleaning the sensors to avoid any
damage to them. You should clean the Spirohome
®
Clinic body between each new patient.
Wipe all accessible surfaces of the device, using moderate pressure, as shown.
⚠
CAUTION:
Care must be taken to prevent any excess liquids contained within the wipes
from entering the components of the Spirohome
®
Clinic. Never immerse the product in water
or any other liquid solution.
3. Perform Disinfection
After cleaning all accessible surfaces of the
device
with
a
high-level
disinfectant
(Sodium
Hypochlorite
<0.6%,
Sodium
hydroxide <0.02%, Sodium Chloride <4%)
wipe, use a second fresh wipe to wipe over
all surfaces again using moderate pressure
and for the contact time recommended by
the
wipe
manufacturer
to
achieve
disinfection.
First Pub. Date: 24.09.2018
R.9-1 / 10.03.2020