Table 2: Immunity (Stimulation mode) table according to IEC 60601-1-2
Guidance and manufacturer’s declaration – electromagnetic immunity
Spirohome
®
battery-operated spirometer devices are intended for use in the
electromagnetic environment specified below. Users of these devices should assure that it
is used in such an environment.
Immunity Test
Standard
IEC 60601 test
level
Compliance
level
Recommended separation
distance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2 kV
±4 kV
±6 kV
±8 kV
±15 kV
±8 kV contact
±2 kV air
±4 kV air
±8 kV air
±15 kV air
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrostatic fast
transient / burst
IEC 61000-4-4
N/A
NA
Surge IEC
61000-4-5
NA
NA
Voltage dips,
short interruptions
and voltage
variations on
power supply
lines IEC
61000-4-11
NA
NA
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical domestic,
commercial or hospital
environment.