36
Parameter
Value
Measure-
ment
accuracy
Oxygen
saturation
±2% (70% to 100%)
No requirement (≤ 69%)
Pulse rate
±2 bpm
Alarm
range
Oxygen
saturation
Upper limit: 50% to 100%
Lower limit: 50% to 100%
Pulse rate
Upper limit: 25 bpm to 250
bpm
Lower limit: 25 bpm to 250
bpm
Alarm
error
Oxygen
saturation
± 1% of the preset value
Pulse rate
The greater of ±10% of the
preset value and ±5 bpm
PI
Weak PI
Min. 0.2%