6. Select the firmware file with the dld extension.
7. The software will now prompt you to put the LM-2 in “Boot Mode.” In order to
accomplish this you must power the unit off. Next, while pressing the
Mode
button on the LM-2, power the unit back ON. The LM-2’s display will show that it
is now in “Boot Mode.” Press the OK button on the software to start the firmware
reflash.
8. Do not disconnect the unit from the computer until the firmware progress screen
completely disappears. Once finished you may disconnect the unit from the
computer and exit out of the software.
Note: Performing a firmware upgrade to the LM-2 clears all calibration data. The
LM-2 will perform a sensor calibration on the next power up after the firmware
upgrade. Be sure to have the O2 sensor outside of the exhaust (free air.)
3.3 Downloading Logs from Memory Card
Note: An SD memory card reader is necessary to retrieve logs from the memory card.
The logs can not be downloaded through the LM-2..
1. Remove the memory card from the LM-2 and insert it in your SD memory card
reader.
2. Launch LogWorks 3. The LogWorks 3 application can be launched from
Start-
>Programs->LogWorks3->LogWorks3
from the Windows task bar.
3. Click on
File->Import/Download->Import LM-2 Log.