15
BUILDER’S GUIDE AND DECLARATION
ELECTROMAGNETIC EMISSIONS
The item INN-028 can be used in the
electromagnetic ambient as it is specified forward.
The client or user of INN-028 must make sure
that the appliance will be used in this of ambient.
Emission test
Compliance
Guide to electromagnetic ambient
Irradiate
Emission/Conducted
CISPR11
Group 1
The product model INN-028 uses RF
energy only for its internal function.
This in because its emissions are very
low and they don’t provoke any in-
terference close to another electrical
apparatus.
Electrostatic discharge
IEC/EN61000-4-2
±6kV contact
±8kV air
Pavements should be of wood,
cement or ceramic. If pavements are
covered by synthetic material, relati-
ve humidity should be max 30%.
Magnetic field
IEC/EN61000-4-8
3A/m
Magnetic field should be typical of a
commercial or an hospital ambient.
Irradiated immunity
IEC/EN61000-4-3
3V/m 80MHz to
2,5GHz
(for no
life-supporting
appliances)
10V/m 80MHz to
2,5GHz
(for no
life-supporting
appliances)