APRIL 2020 • REVISION 4.0
INSTRUCTION MANUAL
37
ISSUES
Issue
On-screen description
Explanation
Possible solutions and advice
16-20
28, 29
There is a hardware issue.
An error has occurred with a
piece of hardware.
For safety, check H
2
O
2
levels with portable gas monitor
before entering room. Please contact Inivos quoting the
issue number to arrange system swap while it is repaired.
23
WiFi communications
issue.
- Something is interfering
with the signal between
the main unit and process
monitor. Please reposition
the process monitor away
from metal and electrical
objects and try again.
Something is interfering with
the communication between
the main unit and the process
monitor. It could be a metal
door, or some electrical
equipment.
Try repositioning the process monitor away from
interfering objects and try again.
If you have run a process successfully in this room before,
please contact Inivos quoting issue 23.
24
Room humidity is too high.
- Please ventilate the room
to reduce humidity. Then re-
seal all vents, windows and
doors and try again.
As this is a vaporising
process, if the ambient room
humidity is too high at the
start, the process cannot run
successfully. Injection has not
started and it is safe to enter
the room.
Turn air conditioning to max. temporarily. Then turn down/
o
ff
again, re-prepare room - re-seal all vents, windows and
doors and run the process again.
On unusually hot and humid days, it may be better to run
the process in the morning or evening.
25
Deactivation fan fault.
- A failure occured during
initial system tests. Please
check H
2
O
2
levels with
portable gas monitor before
entering room to remove
equipment.
Please contact Inivos
quoting issue 25 to arrange
system swap while it is
repaired by us.
A fault has occurred when the
deactivation fans were tested
during the preparation phase.
For safety, check H
2
O
2
levels with portable gas monitor
before entering room.
Please contact Inivos quoting issue 25 to arrange system
swap while it is repaired by us.
26
Injection fan fault.
- A failure occured during
initial system tests. Please
check H
2
O
2
levels with
portable gas monitor before
entering room to remove
equipment.
Please contact Inivos
quoting issue 26 to arrange
system swap while it is
repaired by us.
A fault has occurred with the
injection fans.
For safety, check H
2
O
2
levels with portable gas monitor
before entering room.
Please contact Inivos quoting issue 26 to arrange system
swap while it is repaired by us.
Summary of Contents for ProXcide
Page 8: ...8 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 1 2 ...
Page 41: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 41 Issue 4 Pop up box for issue 4 ...
Page 42: ...42 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 5 Pop up box for issue 5 ...
Page 43: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 43 Issues 12 Pop up box for issue 12 ...
Page 44: ...44 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 13 Pop up box for issue 13 ...
Page 45: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 45 Issues 16 to 20 Pop up box for issues 16 20 ...