It is possible to connect several devices on top of the master device. The maximum
number of devices is shown in the Technical Data. Each device
can be stacked on top of one incubator and connected corresponding to figure 16 →
If you want to connect more devices than recommended for one stack this is possible
but has to be in a second tower. Connections again corresponding to this example
4.6.
Mechanical Integration
The devices can be integrated into a workstation to guarantee a firm stand. Therefore all
devices have position pins.
•
Place the device on an appropriate plate for a firm and even stand (please contact
your workstation provider to find out if they offer an appropriate adapter plate).
• Use the position pins on the bottom side for exact positioning of the device/stack.
Fig.17:
Position Pins
Position pins
NOTE
It is not advisable to stack more than the recommended devices (refer to chapter 2.6) on
top of each other without additional mechanical fixation. There are mechanical limitations
of how many devices can be stacked. Please contact INHECO if you are not sure if your
set up is recommended.
The device has to be in balance for proper shaker performance.
NOTE
If not integrated into a workstation a ground plates with pin holes is also necessary.
Never use the Incubator (especially a shaker) if only placed on pins. This leads to an
unsteady shaking and problems with opening the drawer.