Doc ID: 900434-001
March 2018
INHECO Single Position Incubators
Incubator MP / DWPPart No.: 7300003 / 7300006
Incubator Shaker MP/ Shaker DWPPart No.: 7300013 / 7300009
► User´s Manual
Page 1: ...Doc ID 900434 001 March 2018 INHECO Single Position Incubators Incubator MP DWP Part No 7300003 7300006 Incubator Shaker MP Shaker DWP Part No 7300013 7300009 User s Manual...
Page 2: ...nual This manual and the information herein have been assembled with due diligence INHECO GmbH does not assume liability for any misprints or cases of damage resulting from misprints in this manual If...
Page 3: ...tails General 18 Adress Switch 19 Connecting devices 19 Mechanical Integration 21 Labware Use and Detection 22 Communication with the Incubator 23 Integration in workstation 23 COM port identification...
Page 4: ...s meet the acknowledged rules of technology and comply with today s standards Manual instructions must be followed in order to limit the safety risk during operation of the unit Security related warni...
Page 5: ...roller Microprocessor with on chip peripheral CRC Cyclic Redundany Check FCS Firmware Command Set Heater Same as Temperature Controller IVD D In vitro diagnostic directive MP Micro Plate PCR Polymeras...
Page 6: ...t have been used in Biosafety Level 3 and 4 environments or have been exposed to radioactive or radiation materials Cleaning and Decontamination page 37 Devices exposed to Biosafety level 3 and 4 Envi...
Page 7: ...lling the devices enabling the use of the Incubator products in FDA IVD compliant environments and applications Software and verification tools are available to form a complete thermal solution for al...
Page 8: ...and products and email them to TechHotline inheco com immediately Transportation damages must be reported to INHECO within 7 days of delivery The following components should be included in each shipm...
Page 9: ...ents Figure 2 introduces the terminology of the Incubator s functional elements Fig 2 Functional elements 1 Fan 4 USB Connnector 7 15 pole Sub D Connector male 2 Loading drawer 5 Address Switch 8 15 p...
Page 10: ...section of www inheco com Manual INHECO Incubator Shaker control Incubator MP Incubator DWP 1000 INHECO GmbH 82152 Martinsried Germany 04 PN SN Input 24V 7300003 Incubator MP 2016 09 Rev 30W Fuse F3...
Page 11: ...g Shaker frequency range N A 6 6 Hz to 30 HZ 400 rpm to 1800 rpm Maximum shaking amplitude difference between the maxima N A 3 mm Shaking direction N A Arbitrary programmable e g orgital linear Dimens...
Page 12: ...5 kg Open drawer resistance for a force in Z axis without plastic deformation 10 N Enviromental Conditions Maximum operation height Tolarable relative humidity Operation 10 80 relative non condensing...
Page 13: ...East West Shaker Load 100g Shaker Speed Hz Shaker amplitude mm Single device Maximum stack 10 X 1 0 to 2 2 or Y 1 0 to 2 2 X 1 0 to 1 9 or Y 1 0 to 1 9 15 X 1 0 to 2 7 or Y 1 0 to 2 7 X 1 0 to 2 4 or...
Page 14: ...ture according to a Greiner 96 Flat Bottom Plate Postitions A1 A12 E7 H12 Target Temp Validation Tool Uniformity C 37 C 98 6 F 1 8 Evaporation in unsealed Micro Plates can be reduced by using lid or c...
Page 15: ...nments with risk of explosion The unit is for indoor use only The unit must not be used if the Incubator the power adapter housing or the power cable show visible signs of damage Make sure that the el...
Page 16: ...SD electrostatic discharge sensitive devices Electrostatic charges as high as 4000V accumulate on the human body can discharge without detection Although the Incubator devices feature proprietary ESD...
Page 17: ...the Incubator 4 2 1 Disassembling the transportation screws only for Shaker devices After unpacking the Incubator Shaker device the two red screws transportation lock at the bottom side have to be uns...
Page 18: ...ttom of the device Turn the Incubator Shaker slowly to the left Note Cover side in direction of assembler like in figure 10 unscrew the screw just by hand Fig 9 Screws screwable without screwdriver Af...
Page 19: ...not optimized for a secure real time data transfer all communication is secured by a cyclic checksum CRC In case the communication between the computer and the Incubator Unit often fails due to timeo...
Page 20: ...wers INHECO s compact single position Incubators are high performance plug play devices for the use on automated robotic platforms or as standalone units They can either be used as single devices or s...
Page 21: ...figure 12 on page 21 Slave unit needs no USB cable or power cord Two connected devices 4 5 3 How to connect two devices slave to slave plug in the 15 pole SUB D Connectors The 15 pole SUB D Connector...
Page 22: ...te plate for a firm and even stand please contact your workstation provider to find out if they offer an appropriate adapter plate Use the position pins on the bottom side for exact positioning of the...
Page 23: ...correctly from the device 4 7 1 Incubator MP and Incubator Shaker MP The Incubator MP and Incubator Shaker MP are built for the use with microtiterplates The Labware detection is optimized for Greiner...
Page 24: ...optimized for a secure real time data transfer all communication is secured by a cyclic checksum CRC In case the communication between the computer and the Incubator Unit often fails due to timeouts i...
Page 25: ...hould be found in the list of assigned ports The COM port is shown behind the the USB Serial Port in brackets Please continue with start the workstation software Additional dialogues could appear depe...
Page 26: ...o a robotic systems Please contact the supplier of your robotic system to discuss the integration routine Atypical cycle for a process could contain the following steps Scan COM Ports mandatory Find d...
Page 27: ...r different setup conditions Some devices may be already connected to the USB green LED is on some may not green LED is blinking at 1 Hz The blinking of the LED stops only after the first message with...
Page 28: ...et with your workstation software Manual Workstation Software NOTE Only for Incubator MP device The default Heat up Off set for 37 C is 4 C for a Greiner 96 Flat Bottom Plate If your workstation softw...
Page 29: ...ide gas The surface decontamination can include a wipe down of the housing with a moistened cloth Ethanol 70 can be used where effective against target organisms The use of Bleach 5 12 or Microside SQ...
Page 30: ...O for support INHECO product number of the device shown on device label INHECO product name of the device shown on device label INHECO serial number of the device shown on device label or via software...
Page 31: ...environments are not accepted by INHECO for return Ask techhotline inheco com or visit www inheco com service returns rma html for the return procedure before you return a device to INHECO Do not retu...
Page 32: ...on the lock correctly With this movement the front rear shaker is moved to the back and only then the lock can be placed correctly 2 Turn the unit about 45 60 to the left to position the lock correctl...
Page 33: ...ant check of all data is done The INHECO Measurement Plate IMP can be used for verifation of shaking amplitude and temperature 8 3 Automatically self test AQS The automatically self testAQS which is o...
Page 34: ...contact the support of your workstation provider 9 1 Drawer failure 9 1 1 error 10 check whether the labware matches the labware criteria specified in chapter xxx check whether the opening for the ele...
Page 35: ...re clean check whether the seal ring of drawer is in correct position Fig 29 contact area on drawer lid and correct seal ring position Fig 30 contact pins reboot the incubator 9 1 3 drawer is not clos...
Page 36: ...tails INHECO Part Number Power Supply built for 1 and more incubators AC Input 100 240 V 2 5A 50 60 Hz and DC Output 24 V 7 5 A 180W 2400121 10 2 Power Supply Cord 10 2 1 Order Details INHECO Part Num...
Page 37: ...INHECO 36 Doc ID 900434 001 March 2018 11 Appendix...