Detailed connection description
The VarioCAM
®
/VarioCAM
®
head can be connected with the PC/notebook using the following connection
possibilities. Thereby, it must be noted that certain components are optional and not always supplied as
accessories depending on the configuration of the thermography system.
If the VarioCAM
®
head thermography system is operated without a power supply on a PC
(electricity supply via the FireWire interface), it must be ensured that the FireWire cards
recommended by Infratec are used. Otherwise Infratec makes no guarantee of functionality.
When using the FireWire interface of the VarioCAM
®
/VarioCAM
®
head, the FireWire plug
must be connected to the PC/notebook first and then connected to the VarioCAM
®
with the
LEMO connector.
6.1 VarioCAM
®
head (VCh)
Vch with FireWire cable and power
supply
The VarioCAM
®
head is connected with the 6-pole
LEMO socket using the 6-pole FireWire cable
from the notebook (4-pole)/PC (6-pole). The
power supply is connected to the 14-pole LEMO
socket of the VarioCAM
®
head via the
Adapter P.
The Adapter P (power supply adapter) must
only be used in connection with the power
supply!
Fig. 10 Vch with 6-pole FireWire cable and power
supply
Vch with power supply and universal
cable
The VarioCAM
®
head is supplied with electricity
via the power supply (without adapter). The 6-pole
LEMO connector of the VarioCAM
®
head power
supply is connected with the 6-pole LEMO socket
of the VarioCAM
®
head. The universal cable is
connected to the 14-pole LEMO socket of the
VarioCAM
®
head.
Fig. 11 Vch with power supply and universal cable