Maintenance of the device
12.1
Cleaning
The possible maintenance work to be performed on the VarioCAM
®
HD is limited to the process of
cleaning outer surfaces.
The optical surfaces of the lens are equipped with high-quality optical layers. Avoid any contact
with these surfaces and protect these surfaces against dirt and damages. Therefore, position the
related protective cap on the lens if the device is not in use.
If cleaning the optical surfaces is nevertheless required, please only use wood-free absorbent cotton or
Professional Wipes Kleenex in combination with 96% ethanol. Please contact the manufacturer in the
event of strong contaminations.
When the lens is removed (lens change), the aperture of the camera housing must be protected
against any kind of contact with the inner components. It must be ensured that no dirt can penetrate
the inside of the camera. Interventions of any kind into the inside of the thermography camera are
expressly prohibited.
Maintenance and repair work exceeding the actions and measures described within the framework of this
user manual must only be performed by authorised skilled personnel. If this is not observed, the warranty
and liability obligations of the manufacturer will be forfeit.
12.2
Calibration
The VarioCAM
®
HD system was calibrated very carefully to the official norm with the manufacturer,
maintaining the measure connection in doing so.
We recommend having the calibration checked by the manufacturer or by another certified authority at
intervals of two years. This check should be performed in combination with the general maintenance of
the device with the manufacturer.