background image

• 12 •

• 13 •

Setting Your Meter

The GluNEO

®

 Lite Meter has a wide variety of functions to choose. From the setup mode 

you can turn activity/meal flags, set the date/time, designate three unique averages, and 
set up to five daily alarms.  

Power On

Press        for 3 
seconds at least.

Date / Time

After pressing and 
releasing either 

 or 

buttons to set the date 
and time, respectively 
push       button.  

User Activity Option

After pressing and 
releasing either 

 or 

 

buttons to turn the User 
Activity Option on/off, 
push       button. 

Year

After pressing and 
releasing either 

 or 

 

buttons to set the year, 
push         button.

Number of days for 
average

After pressing and releasing 
either 

 or 

 buttons to set 

the number of days for 
average calculation (3 
different settings possible), 
push       button. 

Alarm on/off

After pressing and 
releasing either 

 or 

 

buttons to turn the alarm 
on/off, push         button.  

User ID Option

After pressing and 
releasing either 

 or 

 

buttons to turn the User 
ID option on/off, push   
button.

Temperature Unit

After pressing and 
releasing either 

 or 

 

buttons to set 

 or 

 

and push       button. 

Alarm time

After pressing and releasing 
either 

 or 

 buttons to set 

the alarm hour and minute, 
push          button. 

Alarm

After programming the 
number of desired alarms 
(5 alarms possible),push    
        button.

CAUTION

1. VERY IMPORTANT: Set the right test unit.
    Your  GluNEO

®

 Lite Meter is pre-set to mg/dL as test unit. A wrong test unit will lead 

to wrong test result display. Please consult your healthcare professional if you’re not 
sure which test unit to use. Without setting the date properly, the average glucose 
level and the results in memory will not show proper values. It is recommended to 
set the glucose test meter before use and also when new batteries are installed.

2. You cannot test your blood glucose in the setting mode.
3. To turn the meter off during the setup, press     for at least 3  seconds.

Test Unit

After pressing and 
releasing either 

 or 

buttons to set the test 
unit(mg/dL or mmol/L), 
push      button. 

Summary of Contents for GluNEO Lite

Page 1: ......

Page 2: ...Blood Glucose Monitoring System for self testing...

Page 3: ...ional before making any changes to your diabetes management The GluNEO Lite Blood Glucose Monitoring System is for in vitro diagnostic use only Important Information The GluNEO Lite Blood Glucose Moni...

Page 4: ...based on the GluNEO Lite meter test results without contacting your physician or healthcare professional first 6 Do not assemble or disassemble the blood glucose test meter 7 Keep out of direct sunli...

Page 5: ...ith GluNEO Lite test strips Do not use during or soon after xylose absorption testing Xylose in the blood will cause interference Before using any product please read all instructions Consult with you...

Page 6: ...Result System Before Having a Meal After Having a Meal After Taking Medication After Sport Activity Control Solution Symbol Blood Drop Symbol for Test User ID Symbol Test Strip Code Symbol Test Result...

Page 7: ...ps in new vials or in any other container 3 Immediately replace the vial cap and close tightly after removing any test strips from the vial 4 Make a note of the expiry date which is six months from th...

Page 8: ...essing and releasing either or buttons to turn the User ID option on off push button Temperature Unit After pressing and releasing either or buttons to set or and push button Alarm time After pressing...

Page 9: ...e able to check that the meter and test strips are working properly and that you are performing the test correctly It is very important that you perform routine checking to ensure accurate results The...

Page 10: ...r animal consumption 6 The control solution bottle icon contains letter C so as not to be confused with the medication bottle This will allow you to understand differences between a control solution t...

Page 11: ...device it provides ultimate safety and comfort for obtaining blood samples An adjustable tip offers 5 different strength levels of skin penetration for individual user s comfort while a lancet ejecto...

Page 12: ...stem Sliding Barrel Release Button Performing a blood test with your fingertip CAUTION 1 If you do not see the code displayed pull the glucose test strip out of the port and re start the procedure fro...

Page 13: ...wist the protective cover off Do not discard To obtain a suitable blood samples Before obtaining a blood sample wash your hands with warm clean water and soap Dry hands completely before testing Step...

Page 14: ...your meter by contaminating it Correct Completely Filled Incorrect Poorly filled Applying Blood Step 4 If you selected to use the User ID option during set up press and button to select the correct U...

Page 15: ...ucose test meter 3 Please follow the waste disposal regulations on the used lancets and strips imposed by your local authorities Reviewing Your Results The GluNEO Lite meter stores up to 365 test resu...

Page 16: ...different areas of your body The figure shown above displays the areas where you can test using your GluNEO Lite meter You may test your blood sugar from your forearm upper arm palm ventral palm or do...

Page 17: ...cemia unawareness To ensure accurate results when lancing your arm forearm or upper arm or palm ventral palm or dorsal hand clean the test site with soap and water Make sure there is no cream or lotio...

Page 18: ...lood glucose test meter Your GluNEO Lite blood glucose test meter is a precision instrument Please handle it with care Lancing device Clean the lancing device and tips with soap and tepid water To dis...

Page 19: ...perature After use close cap tightly Make sure the control solution bottle is tightly closed Record the discard date date vial opened plus 3 months on the control solution on the bottle Do not use if...

Page 20: ...hcare professional NOTE 1 Make sure your date and time are correct after changing your battery 2 Please recycle or dispose of used batteries using your local battery collection systems and in complian...

Page 21: ...movetheBloodglucosetest strip from the GluNEO Lite Blood glucose test meter Dispose of used Blood glucose test strip in a sealed container with your household waste in accordance with local regulation...

Page 22: ...ommenda tions Perform a control solution test to check the accuracy of your test strip Message What it means Action required Trouble shooting Test result is higher than 600 mg dL 33 3mmol L Check the...

Page 23: ...t this instrument will be free from defects in workmanship for 3 years from the date of original purchase Limitations of Warranty This warranty is subject to the following exceptions and limitations 1...

Page 24: ...not Freeze Display Type LCD Reading Time 5 seconds Sample Type Capillary and or venous whole blood Symbol Reference Consult Instructions for use Use By date This product fulfills the requirements of D...

Page 25: ...Memo Memo...

Page 26: ...Memo...

Reviews: