3 7

Technical Specification

3 6

Test Principle and Performance

Sample type

Sample volume

Test Range

Reading time

Calibration

Operating Temperature

Operating Humidity

Strip storage temperature

Display Type

Dimension

Weight 

Power source

Battery Life

Memory Capacity

PC communications port

Capillary whole blood

1.5

10-600 mg/dL (0.6-33.3mmol/L)

9 seconds

Plasma - Equivalent

10-40 °C (50-104

°F

)

10-90%

2-30 °C (35.6-86

°F

)

LCD

77 x 46 x 19 mm

42g (including battery) 

3 V Li Battery (CR2032) x 2

Running 5,000 tests

200 test results
Communication is available to

connect computer with cable

Measurement Range :

The measurement range of the  

finetest Auto-coding

TM 

systems is 10 to 600

mg/dL (0.6 - 33.3 mmol/L).

Test Principle :

Glucose in the blood sample will react to the electrodes in the test strip,

generating an electrical current that will stimulate a chemical reaction. This

reaction is measured by the 

finetest Auto-coding

TM

Meter and displayed as your

blood glucose result.

Note:

Different levels of reactions will occur depending on the amount of

glucose in the blood sample.

Calibration :

The 

finetest Auto-coding

TM

system is plasma-calibrated to allow easy

comparison of results with laboratory methods.

Accuracy :

Studies have shown that the test results of 

finetest Auto-coding

TM 

systems

compare well with laboratory methods.

Precision :

The precision of 

finetest Auto-coding

TM 

System has been assessed by studies in

accordance with ISO 15197.

Results for venous blood samples and Control solutions : 

Blood

43 mg/dL 2,39 mmol/L SD = 1.4

CV = 3.4%

Blood

91 mg/dL 5,05 mmol/L SD = 2.2

CV = 2.4%

Blood

145 mg/dL  8,05 mmol/L SD = 4.2

CV = 2.9%

Blood

223 mg/dL 12,38 mmol/L  SD = 7.7

CV = 3.4%

Blood

360 mg/dL  19,98 mmol/L SD = 6.4

CV = 1.8%

Within Run Precision

Control

41 mg/dL 2,28 mmol/L SD = 1.5 

CV = 3.8 %

Control

113 mg/dL 6,27 mmol/L SD = 3.0

CV = 2.7 %

Control

328 mg/dL 18,20 mmol/L SD = 9.3

CV = 2.9 %

Total Precision

Summary of Contents for Finetest Auto-coding

Page 1: ...Blood Glucose Monitoring System Manual Blood Glucose Monitoring System New bio sensor technology Fast Accurate results in 9 seconds...

Page 2: ......

Page 3: ...Auto codingTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the three year warranty card and send it to us This will ensure that...

Page 4: ...t Auto codingTM Blood Glucose Monitoring System KIT Contents 8 finetest Auto codingTM Blood Glucose Meter 9 finetest Auto codingTM Blood Glucose Meter Display 10 finetest Auto codingTM Blood Glucose T...

Page 5: ...is unopened prior to use and that it contains all of the complete parts shown above If either of these conditions is not met please do not use it Contact the place of purchase to receive a new one 1...

Page 6: ...Opening of the measurement chamber Confirmation window Meter contacting leads Before having a meal After having a meal After taking drugs After sports activity Blood drop symbol for test Test result...

Page 7: ...er the finetest Auto codingTM Strip 7 finetest Auto codingTM test strips are for single use only 8 Keep the test strip bottle away from children 9 Please refer to the user s manual for finetest Auto c...

Page 8: ...ter re setting the test unit See also conversion table on page 38 2 Without setting the date properly the average glucose level and the results in memory will not show proper values It is recommended...

Page 9: ...ough your local representative Control Solution Test can only be used with the finetest Auto codingTM System and should be used during the following cases When a new vial of test strips is opened Any...

Page 10: ...thod is more favorable in preventing contamination to the whole vial of control solution IMPORTANT NOTE If the finetest Auto codingTM Control Solution test results continuously falls outside the range...

Page 11: ...th the code and temperature After 3 seconds the code number and testing temperature will disappear When the strip sign blinks apply your blood to the test port of strip Note Inserting the test strip i...

Page 12: ...the beeping sound the test will begin automatically and your results will appear in 9 seconds It should begin counting down from 9 to 1 second on the LCD display window The LCD window will display th...

Page 13: ...card the used strip and lancet to a proper place according to the local regulation caution 1 If you did not activate the User Activity option and or the User ID option Step 4 1 and or Step 4 2 procedu...

Page 14: ...ID then press the button If you choose the User ID to check the stored results only the results of the selected user are displayed Press shortly If the User ID option has been disabled during setup t...

Page 15: ...e further analyzed You can download for free the finetest Auto codingTM Blood Glucose Program at our web site http www infopia21 com You can purchase the communication cable separately from a local re...

Page 16: ...n if you want to get more information 5 The lancing device should be cleaned if blood or stain remains after to prevent infectious diseases 6 When you need to purchase the lancing device the lancets a...

Page 17: ...em with the Auto coding Label Please insert a new Blood Glucose Test Strip and perform your test again If Er3 message persists please contact your authorized representative or customer support Message...

Page 18: ...r re test with blood sample two or three times When HI message persists please consult a doctor for assistance Problem Caused by What to do Batteries are dead or there is a problem with the meter Chan...

Page 19: ...e a chemical reaction This reaction is measured by the finetest Auto codingTM Meter and displayed as your blood glucose result Note Different levels of reactions will occur depending on the amount of...

Page 20: ...ched to the set Mail or fax it back to us The finetest Auto codingTM Meter has a full three year warranty from the original date of your purchase For your convenience please include your date and plac...

Page 21: ...ive 98 79 EC on in vitro diagnostic medical devices In Vitro Diagnosticum Caution consult accompanying documents Batch code Refrence number Store at Do not re use Use within 3 months apter first openi...

Page 22: ...42 Memo...

Page 23: ...431 080 Korea Authorized representative Obelis S A Bd General Wahis 53 1030 Brussels Belgium Wuxi Xinda Medical Device Co Ltd No 42 Xixin Road Xibei Town Wuxi Jiangsu China Lancing Device Manufacture...

Reviews: