3 7
Technical Specification
3 6
Test Principle and Performance
Sample type
Sample volume
Test Range
Reading time
Calibration
Operating Temperature
Operating Humidity
Strip storage temperature
Display Type
Dimension
Weight
Power source
Battery Life
Memory Capacity
PC communications port
Capillary whole blood
1.5
㎕
10-600 mg/dL (0.6-33.3mmol/L)
9 seconds
Plasma - Equivalent
10-40 °C (50-104
°F
)
10-90%
2-30 °C (35.6-86
°F
)
LCD
77 x 46 x 19 mm
42g (including battery)
3 V Li Battery (CR2032) x 2
Running 5,000 tests
200 test results
Communication is available to
connect computer with cable
Measurement Range :
The measurement range of the
finetest Auto-coding
TM
systems is 10 to 600
mg/dL (0.6 - 33.3 mmol/L).
Test Principle :
Glucose in the blood sample will react to the electrodes in the test strip,
generating an electrical current that will stimulate a chemical reaction. This
reaction is measured by the
finetest Auto-coding
TM
Meter and displayed as your
blood glucose result.
Note:
Different levels of reactions will occur depending on the amount of
glucose in the blood sample.
Calibration :
The
finetest Auto-coding
TM
system is plasma-calibrated to allow easy
comparison of results with laboratory methods.
Accuracy :
Studies have shown that the test results of
finetest Auto-coding
TM
systems
compare well with laboratory methods.
Precision :
The precision of
finetest Auto-coding
TM
System has been assessed by studies in
accordance with ISO 15197.
Results for venous blood samples and Control solutions :
Blood
43 mg/dL 2,39 mmol/L SD = 1.4
CV = 3.4%
Blood
91 mg/dL 5,05 mmol/L SD = 2.2
CV = 2.4%
Blood
145 mg/dL 8,05 mmol/L SD = 4.2
CV = 2.9%
Blood
223 mg/dL 12,38 mmol/L SD = 7.7
CV = 3.4%
Blood
360 mg/dL 19,98 mmol/L SD = 6.4
CV = 1.8%
Within Run Precision
Control
41 mg/dL 2,28 mmol/L SD = 1.5
CV = 3.8 %
Control
113 mg/dL 6,27 mmol/L SD = 3.0
CV = 2.7 %
Control
328 mg/dL 18,20 mmol/L SD = 9.3
CV = 2.9 %
Total Precision