15.1 Status Messages
Some IC6 status messages will have designators to indicate the Material, Source
or Sensor the status refers.
indicates designators and to what they
refer. If there is a message that has multiple occurrences, it will display on one line
with multiple designators after it.
There are a few status messages that have a designated area in the status
message area. These messages are noted in
Table 15-1 Status messages
Status Message
Set
Cleared
Screen
Remote
Comm.
Encode
Crystal Fail X…X
One or more of the sensors
that are called out in currently
active layers fails to measure a
crystal.
X = Sensor
When all sensors in use are
good.
When any sensor in use is
good the sensor number clears
from the numeric list.
All
Designated area
0x8000,0000
0000,0000
Max Power X…X
Whenever the source power
reaches the maximum power
programmed for an active
layer. X = Material number
Whenever the source power is
reduced from the maximum
power for an active layer.
All
Note: This
message will
display in the
cause of stop
area whenever it
is TRUE.
Designated area
0x4000,0000
0000,0000
Test
The general parameter Test
has been set to Yes
The general parameter Test
has been set to No
All.
Designated area
0x2000,0000
0000,0000
A TEST
The general parameter
Advanced Test has been set to
yes.
The general parameter
Advanced Test has been set to
No.
All.
Designated area
0x1000,0000
0000,0000
R Lock
A command was received via
remote communications to lock
the IC6.
NOTE: R Lock supercedes
L Lock)
A command was received via
remote communications to
unlock the IC6.
All
Designated area
0x0800,0000
0000,0000
L Lock
A lock code has been
programmed via the keyboard
and the lock has been set.
NOTE: R Lock supercedes
L Lock)
The lock code has been
entered via the keyboard to
unlock the IC6.
All
Designated area
0x0400,0000
0000,0000
Summary of Contents for IC6
Page 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Page 2: ......
Page 8: ......
Page 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...