6.3.4.2 Setting parameters for “Auto Bar Hole”
1
>
> Auto Bar Hole > Settings (Tab)
2
Measuring Duration. This is the time the detector tests the gas in each
measurement. Select 10 - 20 s (5 s increments)
3
Evacuation Duration. This is the pumping/evacuation time between
measurements. Select 3, 4, 5, 10 or 15 min.
4
CO
2
Limit on Evacuation. This is the limit indicating a well evacuated test point.
Auto Bar Hole sequence will be terminated directly efter completing one
measurement in which CO
2
level has dropped below the set limit. Select Off to
disable function or 1 - 5% to activate function.
Summary of Contents for 580-000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 99: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 99 128 ...
Page 101: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 101 128 ...
Page 102: ...9 Certificates INFICON 102 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 103: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 103 128 ...
Page 105: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 105 128 ...
Page 110: ...9 Certificates INFICON 110 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 111: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 111 128 ...
Page 113: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 113 128 ...
Page 114: ...9 Certificates INFICON 114 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 115: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 115 128 ...
Page 116: ...9 Certificates INFICON 116 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 117: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 117 128 ...
Page 119: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 119 128 ...
Page 120: ...9 Certificates INFICON 120 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 121: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 121 128 ...
Page 122: ...9 Certificates INFICON 122 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 123: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 123 128 ...
Page 126: ...Index INFICON 126 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 127: ......
Page 128: ......