Messages
While operating in the leak detection mode, the display will provide information which
will support the operator using the HLD5000. Besides messages relating to leak
detector settings also warnings and error messages may be displayed.
The HLD5000 is equipped with comprehensive self-diagnostic means. If a fault
condition is detected by the controller board this is indicated via the display to the
operator.
Errors are events which force an interruption to the measurement sequence and
which the HLD5000 is not capable of remedying on its own. Errors are displayed in
plain text on the display. The error sound is a two-tone sound.
The error message can be erased by pressing the "OK” key.
As long as an error status exists, a warning triangle comes up in the lower line of the
display.
An overview over the error messages which may occur while the machine is in use
and possible remedies are listed below.
A list of all error messages as well a notes for troubleshooting is provided in the
Repair Instructions (kipa40e1).
Displayed message
Explanation/possible cause
Remedies
Leak nearly empty
Please replace!
Service life of the Cool Check is less than
3 months
Cool Check is nearly empty
Set on wrong date
Replace Cool Check
Program Main Unit with new date
Probe not connected!
There is no data connection to the sniffer
probe
Sniffer Probe is not connected
If LED at Sniffer Probe is off:
processor in the Sniffer Probe does not
run
Check the sniffer line
Check the LEMO plug or change sniffer
line
Leak temperature is too high!
Main Unit is standing on a hot surface
Ambient temperature too high
Decrease the temperature around the
main unit
Leak temperature too low!
Ambient temperature too low
Increase temperature around the main
unit
Housing temperature too high!
Ambient temperature too high
Fan defect or blocked
Decrease temperature around the main
unit
Change the fan
Flow at measuring line too low!
Measuring line is blocked
Differential sensor is broken
Change the filters of the sniffer tip
Change the CPU-Board (Sniffer Probe)
Leak at measuring line!
Measuring line is leaking
One of the hoses to differential sensor is
leaking
Differential sensor is broken
Check the gasket of sniffer tip
Check or change the hoses to
differential sensor
Change the CPU-Board (Sniffer Probe)
Flow at reference line too low!
Reference line is blocked
Filter is clogged
Differential sensor is broken
Check the sniffer line, change the sniffer
tip
Change the filters of the sniffer tip
Change the CPU-Board (Sniffer Probe)
Summary of Contents for 510-010
Page 2: ...0 2 kina40e IVZ fm technical handbook 1007 ...
Page 8: ...0 8 General Safety Precautions kina40e IVZ fm technical handbook 1007 ...
Page 40: ...4 12 HLD5000 Settings Menu Structure kina40e 04 fm technical handbook 1007 ...
Page 48: ...6 6 Maintenance kina40e 06 fm technical handbook 1007 Fig 6 2 Declaration of Conformity ...
Page 51: ...Appendix A 3 kina40eSIX fm technical handbook 1007 ...