HLD5000 Settings (Menu Structure)
4-7
k
in
a
40
e 0
4
.f
m
te
chn
ica
l ha
nd
boo
k
(1
00
7)
Probe Info 1
•
Internal sensitivity factor S for monitoring the sensor circuitry
•
Differential pressures
Δ
p1and
Δ
p2 to monitor the gas flows in the measurement
and reference lines
•
Channel used by the AD converter
•
Number of operating hours for the probe
•
Software version number for the probe
Probe Info 2
•
HEX Code of probe
•
Contains information for service people
CAL Info
•
CAL factor
•
Phase
•
Sensor factor e.g. R134a
→
R404A
•
Factor which is multiplied with the internal test leak to get the equivalent leak rate
for the measured gas.
Test leak info
•
Leak rate for the calibrated leak normalized to 20 °C
•
Leak rate of the calibrated leak compensated for temperature and type of gas
U / I-Info
•
+ 12 Volt supply voltage (U+)
•
- 12 Volt supply voltage (U-)
•
Supply voltage for the infrared light source in the probe (U light)
•
Current flow through the infrared light source in the probe (I light)
•
Probe test voltage to monitor the voltage supply (U probe)