Prior to installation, store the sensor in the factory supplied container.
Once installed on the sonde, the Temperature Sensor and Conductivity Sensor can be stored wet or dry
depending on the sensor configuration of the sonde. For the best accuracy over instrument life, keep the
conductivity cell submersed in water for 24-48 hours prior to calibration and deployment.
The optical windows should be clear of foreign material. To clear material gently rub the sensing windows
using clean water and a soft cloth or swab. Do not use solvents on the sensor.
Routine Maintenance
Storage
Replacing Wiper Bristles
Replace the Bristles
Replace the Entire Brush
Prior to installation, store the sensor in the factory supplied container. Once installed on the sonde, the
turbidity sensor can be stored wet or dry depending on the sensor configuration of the sonde.
Wiper bristles need to be replaced based on site conditions. In-Situ recommends replacing bristles at least
every 12 months or when visibly bent, damaged, or fouled.
Storage
Cleaning and Storing the Turbidity Sensor
1 Slide both bristle sets out of the wiper head.
2 Insert new small bristle set into the small slot and the large bristle set in the large slot.
1 Loosen the set screw near the base of the wiper head.
2 Slide the wiper head off the wiper shaft.
3 Slide the wipe head on to the wiper shaft, and position the wiper head so the set screw faces the flat part
of the wiper shaft.
4 Tighten the set screw near the base of the wiper head.