In-situ Aqua TROLL 600 Operator'S Manual Download Page 47

47

RDO Salinity Setting

The Aqua TROLL 500 includes automatic salinity compensation. This feature requires a conductivity sensor 

and RDO sensor. With both sensors installed, the sonde uses salinity compensation by default. To change the 

compensation value, follow these steps:

From VuSitu's main 

menu, select 

Connected 

Instrument

.

Select 

Instrument 

Settings

 from the menu 

at the bottom of the 

screen.

From the Instrument 

Settings menu, select 

Salinity Setting

.

Enter your desired 

salinity compensation 

setting and press 

Save

.

1-970-498-1500

www.in-situ.com

Summary of Contents for Aqua TROLL 600

Page 1: ...In Situ In Situ logo Baro Merge BaroTROLL HERMIT HydroVu iSitu Pocket Situ RDO RuggedCable RuggedReader SmarTROLL TROLL VuSitu and Win Situ are trademarks or registered trademarks of In Situ Inc 2016 All rights reserved This product may be covered by patents identified at www in situ com patents ...

Page 2: ...ential Interferents 13 RDO Cap Chemical Incompatibility 13 Conductivity Temperature Sensor 13 Instrument Overview 22 System Components 23 Base Unit Components 23 Getting Started 26 Handling pH and Ion Selective Electrode Sensors ISEs 30 Using the RDO Sensor and RDO Fast Cap 30 Ion Selective Electrode Sensor ISE Requirements 31 Ammonium Sensor Requirements 31 Chloride Nitrate and pH ORP Sensor Requ...

Page 3: ...tes Menu 41 Micro SD Card 42 Replacing Desiccant 44 Sensor Calibration 44 Solution Based Calibration 44 Turbidity Sensor Calibration 45 RDO 100 Saturation Calibration Water Saturated Air 46 RDO 100 Saturation Calibration Saturation Bubbler 46 RDO Salinity Setting 47 Zeroing the Depth Sensor 50 Software 52 VuSitu Mobile App 52 VuSitu Overview 52 VuSitu Menu Options 53 Selecting with Long press and ...

Page 4: ...Serviceable Parts 60 Storage 60 Cleaning the Sonde 61 Cleaning and Storing the pH ORP Sensor 61 Cleaning and Storing the RDO Sensor 63 Cleaning and Storing the Conductivity Sensor 64 Cleaning and Storing the Turbidity Sensor 65 Declaration of Conformity 67 Appendix 68 Appendix A Parameter Discovery 68 Appendix B Parameter Numbers and Locations 69 Appendix C Unit IDs 71 Appendix D Register Data For...

Page 5: ...torage and shipping of your equipment The exclamation point calls your attention to a requirement safety issue or important action that should not be overlooked The Aqua TROLL 600 Sonde is a multiparameter water quality sonde with LCD screen internal memory and additional SD card storage The Aqua TROLL 600 uses the latest sensor and electronics technology to provide laboratory quality measurements...

Page 6: ... received Technical Support will assign an RMA Return Material Authorization number 4 Clean the product as described in the manual 5 If the product contains a removable battery remove and retain it unless you are returning the system for a refund or Technical Support states otherwise 6 Carefully pack your product in its original shipping box if possible 7 Mark the RMA number clearly on the outside...

Page 7: ...ontaminated or if it is the opinion of our Service Representatives that the equipment presents a potential health or biological hazard we reserve the right to withhold service until proper certification is obtained 1 We recommend cleaning with only mild soaps that are compatible with the product materials Wetted materials lists are provided on the product specification sheets Clean all cables and ...

Page 8: ...e instrument Ensure that the RDO Sensor Cap is pressed firmly over the sensor lens and is flush with the instrument before submerging in liquid Replace the cable if insulation or connectors are damaged Make sure the probe and sensor O rings are clean and free of damage Read the safety information on this page before deploying or configuring your Aqua TROLL 600 If you have questions contact In Situ...

Page 9: ... seconds for one parameter no wipe Internal memory 16 MB Additional memory micro SD card2 16 GB included any size micro SD up to 32 GB accepted Logging rate 1 minute to 99 hours Logging modes Linear Linear Average Event Data logging 50 logs defined scheduled to run or stored LCD screen Integrated display shows status of sonde sensor ports data log battery and connectivity Internal power Battery li...

Page 10: ...s a trademark of ExxonMobile Inconel is a trademark of Special Metals Corporation Viton is a registered trademark of DuPont Performance Elastomers L L C 1 For 30 parameters 100 000 data records 3 years at 15 minute interval A single data record includes timestamp temperature RDO pH ORP turbidity and conductivity logged in Linear or Linear Average mode 2 Log data recorded to SD card in comma delimi...

Page 11: ...nt Dimensions with Restrictor Off Total length with connector 60 2 cm 23 7 in Diameter 47 mm 1 85 in Total length with wiper brush on 60 2 cm 23 7 in Total length with wiper brush off 47 mm 1 85 in Diameter 41 66 mm 1 64 in 1 970 498 1500 www in situ com ...

Page 12: ...8 285 9 30 76 200 30 100 250 650 5 to 50 C Barometric Pressure NA 2 years or greater User calibration only if needed NA NA NA 5 to 50 C Ammonium 6 months 6 to 12 months Monthly 30 25 70 0 to 40 C Chloride 6 months 1 year or greater Monthly 350 200 650 0 to 50 C Nitrate 6 months 6 to 12 months Monthly 30 25 70 0 to 40 C Chlorophyll a NA 2 years or greater User calibration only if needed 350 200 650...

Page 13: ...re stronger reducing agents than hydrogen or platinum e g chromium vanadium titanium etc Chloride Hydroxide Ammonia Thiosulfate Bromide Sulfide Iodide Cyanide BGA PC BGA PE Chlorophyll a Rhodamine Turbidity The following chemicals will damage the RDO sensing element Soaking the Conductivity Temperature sensor in vinegar for longer than one hour can cause serious damage Alcohols 5 Hydrogen peroxide...

Page 14: ...2 7 Thallium TI 3100 310 31 pH H pH 1 6 pH 2 6 pH 3 6 Silver Ag 270 000 27 000 2 700 Lithium Li 35 000 3 500 350 Sodium Na 11 100 1 100 110 Ion 100 ppm NH4 10 ppm NH4 1 ppm NH4 Hydroxide OH 3 840 384 38 4 Ammonia NH3 6 0 6 0 06 Thiosulfate S2 O3 2 3 0 3 0 03 Bromide Br 0 68 0 068 6 8 x 10 3 Sulfide S2 9 x 10 5 9 x 10 6 9 x 10 7 Iodide I 1 8 x 10 4 1 8 x 10 5 1 8 x 10 6 Cyanide CN 1 5 x 10 5 1 5 x ...

Page 15: ...si kPa bar mbar mmHg inHg Methodology Silicon strain gauge Accuracy 0 5 of reading plus 1 μS cm from 0 to 100 000 μS cm 1 0 of reading from 100 000 to 200 000 μS cm 2 of reading from 200 000 to 350 000 µS cm Range 0 to 350 000 μS cm Resolution 0 1 µS cm Sensor Type Removeable Response Time T63 1s T90 3s T95 5s Units of Measure Actual conductivity μS cm mS cm Specific conductivity μS cm mS cm Salin...

Page 16: ...Removable with replaceable RDO X Cap Response Time RDO X Cap T63 15s T90 45s T95 60s Units of Measure mg L saturation ppm Methodology EPA approved In Situ Methods under the Alternate Test Procedure process 1002 8 2009 1003 8 2009 1004 8 2009 Accuracy Typical 0 1 full scale FS Range Non vented or Vented 9 0 m 30 ft Burst 27 m 90 ft 30 m 100 ft Burst 40 m 130 ft 76 m 250 ft Burst 107 m 350 ft 200 m ...

Page 17: ...ications pH Sensor Specifications Accuracy 5 0 mV 25 C Range 1 400 mV Resolution 0 1 mV Sensor Type Replaceable pH ORP combo sensor Response Time T63 3s T90 15s T95 30s Units of Measure mV Methodology Std Methods 2580 Accuracy 0 1 pH units or better Range 0 to 14 pH units Resolution 0 01 pH unit Sensor Type Replaceable pH ORP combo sensor Response Time T63 1s T90 2s T95 3s Units of Measure pH unit...

Page 18: ...63 1s T90 1s T95 1s Units of Measure NTU FNU Methodology ISO 7027 Total Suspended Solids Ammonium Sensor Specifications TSS is derived from turbidity Between calibration points User defined reference Range 0 to 1 500 mg L Resolution 0 1 mg L Units of Measure ppt mg L Accuracy 10 or 2 mg L whichever is greater Max Depth 25 m 30 PSI Range 0 10 000 mg L as N Resolution 0 01 mg L Sensor Type Removable...

Page 19: ...l a Sensor Specifications Accuracy 10 or 2 mg L w i g Range 0 190 000 mg L CL Resolution 0 01 mg L Sensor Type Removable Response Time T90 10sec T95 30sec Units of Measure mg L ppm mV Methodology Std Methods 4500 Cl D Accuracy 10 or 2 mg L w i g Max Depth 25 m 30 PSI Range 0 50 000 mg L as N Resolution 0 01 mg L Sensor Type Removable Response Time T90 10sec T95 30sec Units of Measure mg L ppm mV M...

Page 20: ...l dilutions of PC standards from 0 1000 µg L PC Max Depth 200 m Range 0 100 RFU 0 1000 µg L PC Resolution 001 RFU 01 µg L PC Sensor Type Removable Response Time T63 1s T90 1s T95 1s Units of Measure Concentration µg L Fluorescence RFU Excitation Wavelength nominal 590 nm Detection Wavelength 640 nm to 690 nm Linearity R2 0 999 for serial dilutions of PE standards from 0 1000 µg L PE Max Depth 200 ...

Page 21: ...2 0 999 for serial dilutions of RWT Rhodamine Water Tracer standards from 0 1000 µg L Max Depth 200 m Range 0 100 RFU 0 1000 µg L Resolution 001 RFU 01 µg Sensor Type Removable Response Time T63 1s T90 1s T95 1s Units of Measure Concentration µg L Fluorescence RFU Excitation Wavelength nominal 530 nm Detection Wavelength 580 nm to 660 nm 1 970 498 1500 www in situ com ...

Page 22: ...d marine environments It s also ideal for the following applications Groundwater sampling Low flow testing Mine water monitoring Stormwater management The built in LCD screen displays battery status connectivity info and other vital information Setup is easy with the VuSitu app and a Bluetooth enabled mobile device The Aqua TROLL 600 can record data to an internal micro SD card and works with tele...

Page 23: ...or Android 0031240 Mobile Device for Android 0064860 TROLL Com RS 232 Cable Connect 0056140 TROLL Com USB Cable Connect 0052500 TROLL Com RS 232 Direct Connect 0056150 TROLL Com USB Direct Connect 0052510 Cable Stripped and tinned Cable with male connector 0053310 Twist lock Bulkhead Connector 0053240 Twist lock Backshell Hanger Titanium 0051480 Cable Extender 0051490 Large Desiccant titanium conn...

Page 24: ...xygen Calibration Kit 0032110 D O Field Calibration Kit 0080830 Conductivity Calibration Kit Full 0032090 Conductivity Calibration Kit Low 0032630 Conductivity Calibration Kit High 0032640 pH Calibration Kit 0032080 pH ORP Calibration Kit 0032120 pH Storage Solution 0065370 Individual Calibration Solutions See website Ammonium Calibration Kit includes 1 liter each 14 ppm 140 ppm 1400 ppm DI water ...

Page 25: ... vented cable is used with non vented instruments for absolute measurements Compensate absolute measurements by using a BaroTROLL Instrument and Win Situ Baro Merge Software Tefzel vented or thermoplastic polyurethane TPU vented or non vented Cables can be ordered up to 1 219 m 4 000 ft Cables can be ordered with a twist lock termination female connector on both ends that connect to the instrument...

Page 26: ...e supplies 9 Extra sensors 10 D cell alkaline batteries 11 Instrument tools GettingStarted 1 Installthebatteries 2 Installthewipermotorandsensors Open the battery compartment Install alkaline batteries Use Allen wrench to remove and check desiccant color If pink replace Remove restrictor and protective sticker Remove dust cap from wiper motor or motor port plug Apply lubricant to O rings Insert mo...

Page 27: ...ue interlock groove Install remaining sensors 3 Installthewipermotorandsensors Remove protective caps from instrument and cable Ensure O ring on instrument connector is clean Apply vacuum grease to O ring Flat edge inside cable end must align with flat edge on instrument connector Hold textured sleeve of cable in one hand and instrument in other Push and twist until click When using a conductivity...

Page 28: ...or Wireless TROLL Com VuSitu displays the Connected Instrument screen when pairing is complete c d Launch VuSitu and tap Connect The app locates and displays nearby In Situ devices iOS Android An iOS device automatically connects to the closest In Situ instrument To connect to another instrument press Disconnect and then Choose or Add Device VuSitu displays a list of available connections a a b b ...

Page 29: ... situ com 6 Learnmore The product manual provides complete instructions about calibration configuration and deployment See the Frequently Asked Questions sections of the In Situ website for online help https in situ com us faq ...

Page 30: ...yment This step is not necessary for the pH sensor The wiper can severely reduce the life of the RDO Fast Cap Wear will vary by application Verify sensor performance prior to use and replace the Fast Cap if damaged HandlingpHandIon SelectiveElectrodeSensors ISEs UsingtheRDOSensorandRDOFastCap Unscrew reference junction Replace reference filling solution 1 2 Potential salt buildup Rinse with deioni...

Page 31: ...an ISE Ion SelectiveElectrodeSensor ISE Requirements AmmoniumSensorRequirements Chloride Nitrate andpH ORPSensorRequirements To calculate ammonia or or and Conductivity Temperature Sensor Conductivity Temperature Sensor Conductivity Temperature Sensor pH Sensor Temperature Sensor Temperature Sensor 1 970 498 1500 www in situ com ...

Page 32: ...32 SDI 123 Wire 1 970 498 1500 www in situ com ...

Page 33: ...33 ModbusMaster 1 970 498 1500 www in situ com ...

Page 34: ...34 ModbusMasterwithRS232 ConverterRequired 1 970 498 1500 www in situ com ...

Page 35: ...1 or later must be installed on the sonde For information about the full Modbus capabilities of your sonde see the Aqua TROLL 500 600 Interface Specification at www in situ com support type documentation Connect the Twist Lock termination to the instrument and wire the stripped and tinned connection as shown below 1 Install the sensors and turn on the display by holding the instrument vertically a...

Page 36: ...ck is thus 10 1 x 7 5451 5514 The starting register for each parameter points to a block of 7 registers that contain the following information Register Offset Size Registers Mode Access Level R W Data Type Description 0 2 R float The measured value from sensor 2 1 R ushort Data Quality Id If this is 0 then there are no errors or warnings See Full System Specification 3 1 R W float Units Id for the...

Page 37: ...ftware 5 Select the register to read on the PLC using Appendix B a Some PLC devices use the register number directly in programming statements others use register addresses which are one less than the register number the programmer must adhere to the PLC s programming style b Each register is a holding register Some PLCs require you to add 40000 to the register number or address For example 5451 w...

Page 38: ...tuses PossibleConnectedStatuses ErrorMessages PossiblePowerStatuses AccessingtheLCDMenu LCD screen will display sensor status on activation Hold instrument horizontally and slowly tap Aqua TROLL 500 logo 3 4 times to view the main menu Tilt instrument left or right to scroll through menu options Select an item when its background turns black by tapping the instrument once You can enable Bluetooth ...

Page 39: ...nstall sensors Sensor ports are open Install sensors or sensor port plugs Install temperature No Temperature or Conductivity Temperature sensor detected Install a sensor with Temperature Install RDO Cap RDO Cap not detected on RDO sensor Install RDO cap RDO Cap expired RDO cap is no longer valid Install a new RDO Cap RDO Cap XXX days Temporary text message update on the lifespan of the RDO Cap Sta...

Page 40: ... immediately AdjustingtheLCDMenuContrast Activate the LCD menu and scroll to Display Tap once to select Select Contrast Tilt the sonde to the left to decrease the contrast Or tilt the sonde to the right to increase contrast Tap the Aqua TROLL 600 logo to set the contrast a a c c b b d 1 970 498 1500 www in situ com ...

Page 41: ... the firmware Access the language menu Access the updates menu Highlight the desired language and tap Highlight the desired component to update Sonde Sensor 1 2 3 or 4 If the firmware is already up to date the LCD displays message indicating that an update isn t available Select Yes if an update is available The LCD screen confirms the new language a a c c b b 1 970 498 1500 www in situ com ...

Page 42: ...le a log started on November 12 2015 would be named 15111200 CSV 15 is the year 11 is the month 12 is the day and 00 is the log number If multiple logs are recorded on the same day the last number will increase by one for each consecutive log 5 Select the logs you wish to download and move them to the destination 6 Select the logs you wish to delete and press Delete on the keyboard Deleted logs ca...

Page 43: ...e current firmware version on the left then an arrow and the new firmware version on the right 11Scroll to Yes by tilting the instrument Tap the Aqua TROLL 600 logo once 12The LCD screen will display icons while updating Once the update is complete the Aqua TROLL 600 and LCD screen will restart 1 970 498 1500 www in situ com ...

Page 44: ... the inside of restrictor and sensors Follow the instructions in VuSitu to perform the calibration Discard the DI water and repeat rinsing procedure two more times with 10 20 ml of your first calibration standard In VuSitu click Calibrations from the Connected Instrument screen and choose sensor to calibrate Remove the battery cover Remove the wrench from the back of the battery compartment Insert...

Page 45: ... formazine concentration Instead a field will appear in which you can enter the appropriate value Select Set User Defined to begin calibrating with the new value If you are using an In Situ standard and the app does not automatically detect it perform the sensor cleaning and maintenance procedure then select Retry Auto Detect 1 970 498 1500 www in situ com ...

Page 46: ...aturate a small sponge with water Reinstall the end cap and leave sponge in restrictor for five minutes Sponge RDO100 SaturationCalibration SaturationBubbler a d e b c f Fill a 100 saturation bubbler two thirds with tap water Place sonde into bubbler Open the VuSitu mobile app and tap Calibrations RDO Saturation Follow instructions in VuSitu to finish calibration Turn on bubbler and allow 5 10 min...

Page 47: ...he sonde uses salinity compensation by default To change the compensation value follow these steps From VuSitu s main menu select Connected Instrument Select Instrument Settings from the menu at the bottom of the screen From the Instrument Settings menu select Salinity Setting Enter your desired salinity compensation setting and press Save 1 970 498 1500 www in situ com ...

Page 48: ...H 10 to 12 weeks or as required by user protocol or site conditions 12 months Single point Theoretical mV 30 mV 2 or 3 point Slope 66 to 50 mV pH 2 or 3 point Offset 30mV at pH 7 ORP 10 to 12 weeks or as required by user protocol or site conditions 12 months Offset 30 mV RDO 12 months or as required by user protocol 12 months 2 point Slope 0 7 to 1 3 2 point Offset 0 3 mg L Temperature Only when r...

Page 49: ...damine WT Only when required by user protocol 12 months Factory Calibration Factory calibration includes a thorough cleaning full functionality check and sensor adjustments to all applicable sensors over the entire calibrated temperature range We recommend a factory calibration every 12 months or when the unit appears to drift significantly 1 970 498 1500 www in situ com ...

Page 50: ...rtical horizontal angled Do not zero the pressure sensor Erroneous readings will result ZeroingtheDepthSensor Error Code Error Description Recommended Actions UC 0 User Calibration UC Expired 1 Perform a user calibration on the sensor in a known calibration standard 2 Follow individual sensor instructions or user SOPs for calibration FC 0 Factory Calibration FC Expired Call 800 446 7488 or contact...

Page 51: ...ort in situ com ERR 0 Sensor is performing outside of expected range 1 Make sure the sensor is submerged in solution 2 Recalibrate sensor in a fresh batch of calibration standard 3 Reset sensor back to factory defaults and check reading in a calibration standard Recalibrate 4 Call 800 446 7488 or contact In Situ Technical Support support in situ com DIS 0 Sensor is not set up correctly being used ...

Page 52: ...ta to mobile device Transfer data from mobile device to a computer Organize data by Location Calibrate Sensors and View Reports To avoid potential compatibility issues it is important to use the most recent version of the VuSitu Mobile App Find version information and app updates from the Google Play Store or Apple App Store The Aqua TROLL 600 can connect to a Bluetooth enabled device for wireless...

Page 53: ... now select multiple files Press and swipe left to reveal the delete and share icons Press and swipe right to reveal the sharing icon Tap the menu icon in the upper left portion of the screen to view options Tap the icon again to close The features available in the VuSitu mobile app vary slightly depending on the instrument to which it is connected 1 970 498 1500 www in situ com ...

Page 54: ...r On a PC right click on the folder and choose Extract Then open your files in Excel VuSitu displays a progress bar while downloading Choose the data you wish to download To save the entire log to your phone choose All data Select Data Files from the menu at the top left corner of the screen Tap and hold the name of the log you want to share Select Export Choose email cloud storage or another shar...

Page 55: ...ess TROLL Com LiveReadingsinVuSitu SnapshotMode LiveReadingsMode Take a single reading and save to Snapshot file View Snapshot file from Menu Data Files Check Snapshot option View file from Menu Data Files Check Live option Take readings at two second intervals 1 970 498 1500 www in situ com ...

Page 56: ...in any web browser Click the button at the top left to generate a CSV Use the menu at the top left to access the Data Files screen Tap one of the files to view and export Tap Export to save the file and choose how you wish to share it 1 970 498 1500 www in situ com ...

Page 57: ...uLocations HowtoCreateaLocation As an alternative you can manually enter latitude and longitude values and tap Apply Or tap and hold a specific point on the map to drop a pin there Select Locations from the main menu To home in on your mobile device s current location tap the button on the top right Tap the pin icon to establish the location on the map Tap Add Location Enter a name for the locatio...

Page 58: ...ith the Location 1 To Select a Location tap the current Location displayed on the Live Readings screen The list of Locations appears 2 The active Location is marked with a green check mark If no Location has been selected data will be associated with the Default Location 3 Tap the desired location in the list 4 The Live Readings screen appears with the site selected 1 970 498 1500 www in situ com ...

Page 59: ...p the Overflow Menu in the upper right portion of the screen 4 Select Edit Location to make changes or Archive Location to remove it from the list Archived Locations can be restored at any time by tapping the Location accessing the Overflow Menu and tapping Restore Location 1 970 498 1500 www in situ com ...

Page 60: ...s See the sensor instruction sheets for details about maintaining and replacing RDO Sensor Cap Remove the restrictor and the end cap Put the restrictor on the instrument in calibration mode Remove the sensors and wiper Dampen the sponge inside the pH sensor storage cap with Storage Solution or pH calibration standard Replace the caps at both ends of the sensor Use electrical tape to seal the stora...

Page 61: ...ife of the sensor The sensor fill solution has a shelf life of 2 years Replace the fill solution every 5 to 6 months or when 1 The sensor fails to calibrate within the acceptable slope and offset range 2 Sensor readings vary 3 Readings during calibration at pH 7 are greater than 30 mV or less than 30 mV 4 Sensor is slow to respond If the sensor fails to calibrate after you replace the fill solutio...

Page 62: ...method and continue to the others only if necessary Do not directly wipe the glass bulb To clean the pH sensor gently rinse with cold water If further cleaning is required consider the nature of the debris To remove crystalline deposits 1 Clean the sensor with warm water and mild soap 2 Soak the sensor in 5 HCl solution for 10 to 30 minutes 3 If deposits persist alternate soaking in 5 HCl and 5 Na...

Page 63: ...er 3 Gently wipe with a soft cloth or brush if biofouling is present 4 If extensive fouling or mineral buildup is present soak the sensor in vinegar for 15 minutes then soak in deionized water for 15 minutes Routine Maintenance Do not use organic solvents they will damage the sensor cap Do not remove the sensor cap when rinsing or brushing 5 After cleaning the sensor perform a 2 point calibration ...

Page 64: ... a soft brush to gently clean the sensor pins and temperature button Ensure removal of all debris around the base of the pins and button 3 If crystalline deposits persist soak in 5 HCl for 10 to 30 minutes followed by warm soapy water and soft brushing 4 If deposits persist alternate soaking in 5 HCl and 5 NaOH solutions followed by warm soapy water and soft brushing Clean the sensor face with war...

Page 65: ... installation store the sensor in the factory supplied container Once installed on the sonde the turbidity sensor can be stored wet or dry depending on the sensor configuration of the sonde Wiper bristles need to be replaced based on site conditions In Situ recommends replacing bristles at least every 12 months or when visibly bent damaged or fouled Storage CleaningandStoringtheTurbiditySensor 1 S...

Page 66: ...ze and its biocidal properties decrease Cleaning the copper restores the effectiveness of the restrictor s antifouling capability 1 Remove the restrictor from the sonde 2 Remove the restrictor end cap 3 Submerge the restrictor in white vinegar acetic acid for 24 hours 4 Rinse the restrictor in water and air dry 66 1 970 498 1500 www in situ com ...

Page 67: ...67 DeclarationofConformity 1 970 498 1500 www in situ com ...

Page 68: ...ontents of these registers indicate which parameter IDs 1 to 219 are currently available from the sonde according to the table below Refer to Appendix B for a description of parameter IDs Parameter ID Map Register Bit 15 14 13 2 1 0 6984 16 15 14 3 2 1 6985 32 31 30 19 18 17 6986 48 47 46 35 33 33 6987 64 63 62 51 49 49 6988 80 79 78 67 65 65 6989 96 95 94 83 81 81 6990 112 111 110 99 97 97 6991 1...

Page 69: ...3 65 uS cm 11 Resistivity 5521 5520 81 ohm cm 12 Salinity 5528 5527 97 PSU 13 Total Dissolved Solids 5535 5534 114 ppt 14 Density of Water 5542 5541 129 g cm3 16 Barometric Pressure 5556 5555 22 mmHg 17 pH 5563 5562 145 pH 18 pH mV 5570 5569 162 mV 19 ORP 5577 5576 162 mV 20 Dissolved Oxygen Concentration 5584 5583 117 mg L 21 Dissolved Oxygen Saturation 591 5590 177 Saturation 24 Chloride Cl 5612...

Page 70: ...n NH4 N concentration 5717 5716 117 mg L 40 Ammonium NH4 mV 5724 5723 162 mV 41 Ammonia as Nitrogen NH3 N concentration 5731 5730 117 mg L 42 Total Ammonia as Nitrogen NH3 N concentration 5738 5737 117 mg L 48 Eh 5780 5779 162 mV 49 Velocity 5787 5786 118 µg L 50 Chlorophyll a Concentration 5794 5793 118 µg L 51 Chlorophyll a Fluorescence Intensity 5801 5800 257 RFU 54 Blue Green Algae Phycocyanin...

Page 71: ...ar millibars 22 mmHg Millimeters of Mercury 0 to C 23 inHg Inches of Mercury 4 to C 24 cmH2 O Centimeters of water 4 to C 25 inH2 O Inches of water 4 to C 26 Torr Torr 27 atm standard atmosphere Distance Length 33 48 33 mm millimeters 34 cm Centimeters 35 m Meters 36 km Kilometer 37 in Inches 38 ft Feet Coordinates 49 64 49 deg Degrees 50 min Minutes 51 sec Seconds 1 970 498 1500 www in situ com ...

Page 72: ...milligrams per liter 118 µg L micrograms per liter 119 deprecated no longer available was µM L was used but invalid 120 g L grams per liter 121 ppb Parts per billion Density 129 144 129 g cm3 Grams per cubic centimeter pH 145 160 145 pH pH Voltage 161 176 161 µV micro Volts 162 mV milli Volts 163 V Volts Dissolved Oxygen DO Saturation 177 192 177 sat Percent saturation Turbidity 193 208 193 FNU fo...

Page 73: ... gallons per day 217 m3 sec Cubic meters per second 218 Available was Cubic meters per minute 219 m3 hr Cubic meters per hour 220 Available was Cubic meters per day 221 L s Liters per second 222 ML day Millions of liters per day 223 mL min milliliters per minute 224 kL day Thousands of liters per day Volume 225 240 225 ft3 Cubic feet 226 gal Gallons 227 Mgal Millions of gallons 228 m3 Cubic meters...

Page 74: ...eet per second 290 mA Milli Amps 291 A Amps Velocity 305 320 305 ft s Feet per second 306 s Meters per second AppendixD RegisterDataFormats The Modbus protocol specification requires any multiple byte data type to be transmitted in Big Endean order or most significant byte MSB first In Situ devices shall use the following register data formats ID Type Size registers Description 2 Unsigned Short 1 ...

Reviews: