Touching the screen with two fingers 
simultaneously will freeze the image.

Touching the screen with one finger  
will open the main menu.

4.2 Main menu

Primary 

device 

indicator

Meat quality

Adjust scanning depth

Frame 

averaging

Start recording

Open Settings
Adjust gain

Freeze ultrasound

To enable frame averaging, touch the icon. To change the number of frames being averaged, 
touch the icon (each touch increases the number of frames up to 4 then returns to 2). To disable 
frame averaging, long press the icon.
Frame averaging combines the current frame with the previous 2, 3 or 4 frames, this produces a 
smoother image with less noise, but at the cost of blurring when the probe is moving quickly.

Summary of Contents for Easi-Scan: Go

Page 1: ...TM Your complete animal imaging solution Instruction manual...

Page 2: ...expressly approved by the party responsible for compliance it could void the user s authority to operate the equipment This device complies with Industry Canada s licence exempt RSSs Operation is sub...

Page 3: ...ress to change channel when no viewing devices are connected Up button Adjust gain Save videos Down button Adjust depth Save images 1 1 Scanner menu In live scanning pressing Up button will bring up t...

Page 4: ...the data while scanning It is recommended that you save images and videos directly from the app where possible The cine loop must be downloaded by the app before switching off otherwise it will be lo...

Page 5: ...pp retrieves the license from IMV s license server and the app will request to connect back to the scanner 6 Select your EASISCAN from the Wi Fi page and return to the app After a few seconds the ultr...

Page 6: ...inished the scanner will switch off The scanner software version number can be checked in the User Settings section 4 IMV Go Scan App 4 1 Live scanning Change the channel in Wi Fi settings No indicato...

Page 7: ...ng Start recording Open Settings Adjust gain Freeze ultrasound To enable frame averaging touch the icon To change the number of frames being averaged touch the icon each touch increases the number of...

Page 8: ...ultrasound Open Settings Adjust gain Scan Go Manual full Eng Ver AW V27 Sep 2018 www photodesign studio com indd 7 27 9 18 23 40 Go to cine mode Callipers to measure size of the embryo Save image on...

Page 9: ...23 40 Crown rump length Trunk diameter Eye diameter Head diameter Head length Callipers TM 8 Callipers Reset the callipers to the initial position Save images to the camera folder on a smart device Re...

Page 10: ...surements available Crown Rump length distance from the top of the skull to the end of the body Trunk diameter width of the ribcage at the widest part Eye diameter diameter of the eye socket Head diam...

Page 11: ...he body 14mm 180mm 33 76 days distance from the top of the skull to the end of the nose 21mm 146mm 56 134 days Scan Go Manual full Eng Ver AW V27 Sep 2018 www photodesign studio com indd 9 27 9 18 23...

Page 12: ...button deletes all drawn lines Save button save the current image to the camera folder Back button return to freeze menu Frame back Frame forward Back Frame counter Save cine Save image Play Frame sli...

Page 13: ...the sensitivity at minimum the app looks for the strongest reflection with the sensitivity at maximum it uses a mixture of the average value and the maximum which can compensate for a poor or noisy i...

Page 14: ...If the app considers the measurement to be bad either line is red then you cannot save the record Touch the meat quality icon here you can enter the animal I D and the weight of the animal Touch the f...

Page 15: ...next measurement you save will prompt for a new file name Each entry in the file will contain the animal id weight fat thickness meat thickness date and time and the image reference for this measureme...

Page 16: ...Guide FAQ BUGs brightness BUGs orientation Adjust screen brightness Restores to original settings when the app is closed Adjust ultrasound contrast Select probe orientation The probe orientation will...

Page 17: ...icence expiry date Scanner serial number Software version of app Number of usage files on this device when it was last synchronised with the server Practice or company name User email User name Yellow...

Page 18: ...com Automatically if launching app when connected to Wi Fi Manually by clicking sync button Prompts If there are more than 20 usage files to upload Our privacy policy can be viewed on our website or...

Page 19: ...ower pack is through the gold pads on battery and scanner Power pack output is protected but care should be taken to prevent short circuiting the gold connector pins and pads LED sequence One blue fla...

Page 20: ...pe reducing the need for drying This prevents problems relating to corrosion following wash down The scanner and power pack can also be cleaned using a cloth soap and warm water Sponging down or even...

Page 21: ...ging from a vehicle supply Only replace fuse with 2A medium acting fuse rated to 125V 10kA breaking capacity F2A M125V DC power Adapter XP Power Model VEP24US12 Input 100 240VAC 0 6A Output 12VDC 2 0A...

Page 22: ...thoroughly check the instrument and advise of any work that appears to be necessary If your Easi Scan requires servicing please contact an official authorised service center or the distributor in your...

Page 23: ...ce IMV imaging France 126 Boulevard de la R publique 16000 Angoul me France 33 5 45 92 03 57 france imv imaging com www imv imaging fr India IMV India Pvt Ltd Plot No 750 Phase V Udyog Vihar Gurugram...

Page 24: ...plete animal imaging solution facebook com IMVimaging twitter IMVimaging www imv imaging com info imv imaging com 44 0 1506 460023 Contact us now IMV imaging V 1 February 2019 Instruction manual Easi...

Reviews: