background image

22

  |  

Size – 

 532mm(L) x 67.5mm(W) x 134mm(H)

Weight – 

1kg

Mechanical - Aluminium with PC/ABS mouldings 

and PMMA specula.

AlphaVision operating frequency bands: 

B1=2412MHz-2462MHz and B2=5150MHz-5250MHz. 

Maximum radio-frequency power transmitted in 

the frequency bands: 16.1dBm for B1 and 18dBm 

for B2. 
To comply with ISED RSS-247, 5150-5250MHz bands 

are for indoor use only. 

User Output

AlphaVision wireless link to compatible viewing 

device using IMV Sound-Link over standard WI-FI 

connection.

Image and video storage

Storage of images and video clips limited only 

by available space on smart device linked with 

AlphaVision.

Smart devices compatibility

Android:

Minimum version: Kit Kat (Android V4.4) Minimum 

RAM: 2Gb

Minimum Processor: Quad-Core 1.8GHz 

Recommended WiFi: 802.11 a/ac (5GHz band)

Apple:

Minimum version: iOS 8

Minimum iPhone version: iPhone 5s Minimum iPad 

version: iPad4 

Specifications

Power and charging  

Removable li-ion battery gives 7h runtime.
Battery charge time 5.5h. Only use 

IMV imaging

 

approved/supplied battery, 12V vehicle power 

supply, battery charger and DC power adapter.
ESG-CHARGER: Input: 12V, 2A / Output: 4.2V, 2A.

IMV imaging

 Vehicle Power Outlet/Cigarette 

Plug Cable (Model: ACC-DC LEAD):12VDC, 2A 

Fused.
Only to be used when charging from a vehicle 

supply.
Only replace fuse with 2A medium acting fuse 

rated to 125V, 10kA breaking capacity (F2A 

M125V).
DC power Adapter (XP Power, Model: VEP24US12): 

Input: 100-240VAC, 0.6A/Output: 12VDC, 2.0A
Battery (Model: ESG-BATT): 3.6Vdc, 6700mAhr.

For further information and video manuals about 

the AlphaVision please visit the “Customer 

support“ section of the IMV website 

www.imv-

imaging.com/international/service-support.

 

Temperature range:
 

AlphaVision: -10°C to 45°C *

 

Battery Charger: 0°C to 35°C *

Relative humdity: 50% RH max.
Maximum use altitude < 2000m.
Pollution degree: 2
Battery Charger has under and over temperature 

protection and LED status indication.
*

 Best charged at 20-25°C, frequent charging at       

35 °C / 95 F may shorten battery lifetime.

 

Environmental 

conditions

IMV imaging

 warrants AlphaVision against 

defects in materials for one year from the date of 

purchase. 

IMV imaging

 does not warrant against 

normal wear and tear, nor damage caused by 

accident or abuse. 

Warranty

!

Warning: 

if the equipment is used in a  

manner not specified by the 

manufacturer, the protection provided 

by the equipment may be impaired.

Summary of Contents for AlphaVision

Page 1: ...Your complete animal imaging solution Instruction manual...

Page 2: ...2...

Page 3: ...n 4 Preamble 5 Introduction 6 Overview 7 Carry case composition 8 Activation 9 Operating instructions 13 Dismantling the device 20 Care and maintenance 21 Cleaning 21 Specifications 22 Warranty 22 Env...

Page 4: ...ull text of the EU declaration of conformity is available at the following internet address Certification www imv imaging com avv2 doc Ag ncia Nacional de Telecomunica es Este equipamento nao tem o di...

Page 5: ...ere procedures and specifications have not been correctly followed It is therefore essential that the user follows the specifications procedures and recommendations described in this user manual Befor...

Page 6: ...insemination procedure less invasive it also helps to investigate cervix malformation IMV imaging has utilised the latest camera and WiFi technology in AlphaVision to deliver the highest quality imag...

Page 7: ...www imv imaging com 7 Overview SPECULUM ASSY SOFT TIP MEC09 013 SPECULUM ASSY NO TIP MEC09 014 CHASSIS HANDLE MEC09 002 BATTERY ESG BATT KOMBICOLOR GUIDE MEC09 015 CAMERA ALPHAVISION SYSTEM...

Page 8: ...SG CHARGER 1 bottle of B LUBE gel ref 023917 1 straw cutter ref 005233 2 Kombicolor insemination guns ref 018398 018400 1 Treatment sheath 50 Alpha sheaths ref 024485 1 Carry case ref PAK15041 150 san...

Page 9: ...t up process Your AlphaVision can be connected to Android and Apple devices To find the app search for IMV AlphaVision WiFi in either the Google Play store or the Apple store and install the app Lanch...

Page 10: ...ds the Save Video pop up will appear Save Video pop up This menu appears when recording stops the meta data is optional if entered it is stored alongside the video as a text file Selecting Yes will sa...

Page 11: ...g from the list Any meta data associated with the file will be displayed If text is typed in the search box only images and videos which contain that text will be visible in the list The app searches...

Page 12: ...a different folder IMV_AlphaVision_Uploaded and are no longer visible in the Review data menu System update The app automatically checks the IMV licence server for scanner software updates when conne...

Page 13: ...utton Buttons The AlphaVision has 2 buttons on the keymat Power button one long press 2 3 seconds will turn the device on Repeat this to turn the device off Image and video capture button A single pre...

Page 14: ...y and the AlphaVision Power pack output is protected but care should be taken to prevent short circuiting the gold connector pins and pads Operating instructions Warning To reduce the risk of burns fi...

Page 15: ...d feel secure Operator safety When using viewing devices in combination with your Alphavision and working with animals please be aware of your own spatial environment to avoid colliding with unseen ob...

Page 16: ...ot wind the sheath around the speculum this will trap the sheath and make it difficult to open 1 Always work with a dry speculum Wipe carefully with a paper towel between uses 2 Pull the sanitary shea...

Page 17: ...the camera upwards at a 45 angle 4 Gently insert the speculum aligning it parallel to the vagina Operating instructions 5 Once it is 10cm inside the vagina pull back on the sanitary sheath to open it...

Page 18: ...sheath into the Kombicolor until it is fully locked 4 Click the Kombicolor gun into the Kombicolor guide 5 Slide the Kombicolor guide into the AlphaVision handle up to the second notch 1 After thawing...

Page 19: ...olor guide 1 notch or click is equal to 1cm 5 Check that the end of the insemination syringe is positioned correctly by rectal palpation If the tip of the sheath is still in the cervix move the cervix...

Page 20: ...rwards so as not to soil the inside of the AlphaVision 2 Remove the Kombicolor guide from the handle and pop the Kombicolour out of the guide 3 Remove the speculum by pressing the two spring plungers...

Page 21: ...t be removed by rinsing gently do not rub the lens The AlphaVision and battery can be wiped down with an antibacterial wet wipe reducing the need for drying This prevents problems relating to corrosio...

Page 22: ...er adapter ESG CHARGER Input 12V 2A Output 4 2V 2A IMV imaging Vehicle Power Outlet Cigarette Plug Cable Model ACC DC LEAD 12VDC 2A Fused Only to be used when charging from a vehicle supply Only repla...

Page 23: ...Meath K32 ER81 Ireland 353 0 42 932 0070 ireland imv imaging com www imv imaging ie United Kingdom HQ IMV imaging Imaging House Phoenix Crescent Strathclyde Business Park Bellshill ML4 3NJ Scotland UK...

Page 24: ...imaging solution facebook com IMVimaging twitter IMVimaging www imv imaging com www alphavision imv com info imv imaging com 44 0 1506 460023 Contact us now IMV imaging V 1 April 2022 Instruction man...

Reviews: