Start-up
13
5
IMT Analytics AG
2. Screw the oxygen sensor clockwise into the appropriate port. Make sure the sen-
sor seals off the port and there is no leak.
Figure 5: Screwing in the oxygen sensor
3. Connect the cable included to the oxygen sensor by pushing the cable into the hole
at the top of the sensor until the cable locks into place. Connect the other end of
the cable to CITREX H3 by inserting it into the hole provided, which is labelled "O
2
".
Figure 6: Oxygen sensor cable
4. Perform an oxygen calibration. The calibration procedure is described in the sec-
tion "Calibration". Calibration ensures that the measured values of the new sensor
are correct.
Summary of Contents for CITREX H3
Page 1: ...User Manual CITREX H3 ...
Page 2: ...IMT Analytics AG Gewerbestrasse 8 9470 Buchs SG Switzerland www imtanalytics com ...
Page 44: ......
Page 45: ......