15.4 Gas standards for flow and volume readings
CITREX H3 converts the flow and volume readings measured in the device to match the conditions of the standard selected. The
following gas standards are supported by CITREX H3.
Gas Standard
Abbrevia-
tion
Pressure
Temperature
Relative
humidity
Ambient Temperature and Pressure
ATP
Current ambient pressure
Current
gas temperature
Current
gas humidity
Ambient Temperature and Pressure Dry
ATPD
Current ambient pressure
Current
gas temperature
0%
Ambient Temperature and Pressure Saturated
ATPS
Current ambient pressure
Current
gas temperature
100%
Ambient Pressure at 21°C
AP21
Current ambient pressure
21.0 °C (70 °F)
Current
gas humidity
Standard Conditions USA
STP
1013.25 mbar (760 mmHg)
21.0 °C (70 °F)
0%
Standard Conditions USA Humid
STPH
1013.25 mbar (760 mmHg)
21.0 °C (70 °F)
Current
gas humidity
Body Temperature and Pressure, Saturated
BTPS
Current ambient pressure
+ channel pressure
37.0 °C (99 °F)
100%
Body Temperature and (Ambient) Pressure
Saturated according to ISO 80601-2-12:2011
BTPS-A
Current ambient pressure
37.0 °C (99 °F)
100%
Body Temperature and Pressure Dry
BTPD
Current ambient pressure
+ channel pressure
37.0 °C (99 °F)
0%
Table 13: Gas standards for flow and volume readings