Start-up
12
5
IMT Analytics AG
5.2.2
Oxygen sensor (option)
CITREX H3 can be fitted with an oxygen sensor. The option is enabled using the con-
figuration tool. You will find further information on the configuration tool in the section
"Configuration tool".
After enabling, oxygen concentration can be measured in the flow channel. To do so,
an oxygen sensor is screwed into the appropriate port. The oxygen sensor has to be
connected to the measuring instrument using the cable included. The following steps
explain how to install and replace the oxygen sensor.
Oxygen sensor holder
Figure 3: Oxygen sensor holder
Measuring range 0 – 100%
Accuracy
± 1% O
2
(absolute)
5.2.3
Installing the oxygen sensor
1. Remove the protective cap from the sensor port of the device.
Figure 4: Protective cap