CRYOSAUNA TROUBLESHOOTING GUIDE 

Summary of Contents for CRYOSAUNA

Page 1: ...CRYOSAUNA TROUBLESHOOTING GUIDE...

Page 2: ...RGEE134 Settings for the temperature input s measurement object are incorrect or the measurement object failed DO NOT run any Cryosauna Sessions Check RTD and cable connections for tightness If the qu...

Page 3: ...sses thermal insulating gloves long sleeves long pants and closed toed shoes then close the liquid valve on the Dewar For a bulk tank don safety glasses thermal insulating gloves long sleeves long pan...

Page 4: ...ng due to the liquid in the tank converting to gas The LN2 Dewars have an overpressure regulator set to 28 psi This is a normal occurrence and is a sign the LN2 Dewar is maintaining pressure at or nea...

Page 5: ...ave a portable Oxygen Monitor change the battery It is a good idea to have extra batteries on hand If the Oxygen Monitor is malfunctioning contact Impact Cryotherapy support 13 There is a popping nois...

Page 6: ...iple times during the day Pressure drops on the Dewar after multiple Sessions have been run Either open the pressure building valve fully until the valve begins to frost or slightly open the pressure...

Page 7: ...that it is in the ON position See picture below If all the above check out and the Cryosauna still will not turn on please contact Impact Cryotherapy support 20 The stairs are no longer sturdy or the...

Page 8: ...auna is ready for use To address the LOW PRESSURE message see 16 of the Troubleshooting Guide 23 Cryosauna won t run Session or Pre Cool Required The pressure in the Cryosauna Nitrogen Delivery System...

Page 9: ...rt impactcryo com Sales sales impactcryo com Corporate Phone Line 866 335 7768 Web www impactcryo com CRYOSAUNA INFORMATION Cryosauna S N _________________________ Cryosauna Year ________________ Cryo...