CRYOSAUNA TROUBLESHOOTING GUIDE 

If the door latch tension is too low, use a 
flathead screwdriver and turn the 
adjustment screws to the right to 
increase the tension.  

 

19 

The Cryosauna 

won’t start.

 

Check the power 
supply. 

Confirm the Cryosauna is plugged in. 

If you are using a surge protector, 
confirm it is ON. 

Check the outlet for power. 

Check the handle on the Electrical 
Control Panel to make sure that it is in 

the “ON” position. See picture below.

 

If all the above check out and the 
Cryosauna still will not turn on, please 
contact Impact Cryotherapy support. 

20 

The stairs are no longer sturdy, or the stair 
rail is loose. 

The stair placement is 
important to maintain 
safe entry and exit. 

Check the placement of the stairs and 
re-align them to the front of the unit. 

If the stair rail has come loose, tighten it 
with an Allen wrench or hex key. 

21 

Cryosauna 

won’t run Session…Pre

-Cool 

Required 

 

The temperature in 
the Cryosauna is 
warmer than -40° C or 
-40° F 

No one is allowed inside the 
Cryosauna during the Pre-
Cool Cycle.  

Pre-Cool Cycle is required & entry is 

prohibited.  

Also note the “LOW 

PRESSURE” message, see #16 of the 

Troubleshooting Guide. 

 

 

 

 

 

 

 

 

 

Summary of Contents for CRYOSAUNA

Page 1: ...CRYOSAUNA TROUBLESHOOTING GUIDE...

Page 2: ...RGEE134 Settings for the temperature input s measurement object are incorrect or the measurement object failed DO NOT run any Cryosauna Sessions Check RTD and cable connections for tightness If the qu...

Page 3: ...sses thermal insulating gloves long sleeves long pants and closed toed shoes then close the liquid valve on the Dewar For a bulk tank don safety glasses thermal insulating gloves long sleeves long pan...

Page 4: ...ng due to the liquid in the tank converting to gas The LN2 Dewars have an overpressure regulator set to 28 psi This is a normal occurrence and is a sign the LN2 Dewar is maintaining pressure at or nea...

Page 5: ...ave a portable Oxygen Monitor change the battery It is a good idea to have extra batteries on hand If the Oxygen Monitor is malfunctioning contact Impact Cryotherapy support 13 There is a popping nois...

Page 6: ...iple times during the day Pressure drops on the Dewar after multiple Sessions have been run Either open the pressure building valve fully until the valve begins to frost or slightly open the pressure...

Page 7: ...that it is in the ON position See picture below If all the above check out and the Cryosauna still will not turn on please contact Impact Cryotherapy support 20 The stairs are no longer sturdy or the...

Page 8: ...auna is ready for use To address the LOW PRESSURE message see 16 of the Troubleshooting Guide 23 Cryosauna won t run Session or Pre Cool Required The pressure in the Cryosauna Nitrogen Delivery System...

Page 9: ...rt impactcryo com Sales sales impactcryo com Corporate Phone Line 866 335 7768 Web www impactcryo com CRYOSAUNA INFORMATION Cryosauna S N _________________________ Cryosauna Year ________________ Cryo...

Reviews: